Capnography Monitoring


Figure 8-1 Infrared (IR) light absorption by the carbon dioxide (CO2) molecules.


Sidestream Capnography


Sidestream capnography (Figure 8-2) is the most widely used method for continuous CO2 monitoring in the operating room. It involves the use of disposable tubing (6 feet long) and a T-piece adapter, which is inserted between the breathing circuit and the endotracheal tube or other airway device. The tubing is connected from the side port on the adapter to a separate monitor unit, which contains the CO2 sensor. A sample of gas is aspirated through this disposable tubing during the respiratory cycle into the capnometer for measurement. Because this gas sample must travel through the tubing to the CO2 sensor before it is processed, a slight delay occurs in the display of the CO2 waveform. One of the main advantages of sidestream capnography is that it can be used in nonintubated patients. For example, modified nasal cannulas allow the sampling of expired gases even while supplemental oxygen is administered. A drawback of sidestream capnography is that the tubing may become blocked by water vapor or secretions. The use of a filter between the tubing and the unit containing the CO2 sensor minimizes this problem. Keeping the sample tubing antigravity may also minimize contamination from water vapor or secretions.2


f08-02-9780323085120

Figure 8-2 Sidestream Capnography.

The infrared (IR) light sensor is contained within the capnometer, not in the airway circuit.

Mainstream Capnography


Mainstream capnography (Figure 8-3) involves the use of an adapter between the breathing circuit and the endotracheal tube. In this method, a lightweight infrared sensor is attached directly to the adapter. The sensor emits infrared light to a photodetector on the other side of the adapter. During the respiratory cycle, respiratory gases flow through the adapter, and the amount of CO2 in the sample is measured immediately. No extra tubing is required, and no delay occurs in the display of the waveform. The mainstream sensor is heated above body temperature, which prevents condensation and allows the sensor to function in high-moisture environments. Condensation of moisture may interfere with the functioning of the unit. The earlier generations of mainstream capnometers had significant disadvantages, including bulky sensors requiring sterilization after each use and a risk of facial burns from the heated sensor. Newer versions, however, are more lightweight and have disposable adaptors. The temperature of the sensors is lower with better shielding to minimize the risk of facial burns. Further advances in mainstream capnography have produced smaller lighter weight adapters for use in nonintubated patients also. These units attach directly to the oxygen facemask or nasal cannula, allowing for CO2 monitoring in a spontaneously breathing patient receiving sedation.2


f08-03-9780323085120

Figure 8-3 Mainstream Sensor Capnography.

The infrared (IR) light sensor is positioned within the airway circuit.

Colorimetric Devices


Colorimetric devices are portable, disposable, single-use devices that contain a pH-sensitive chemical indicator, which changes color on exposure to CO2. Similar to the mainstream capnometer placement, this device is connected between the circuit and the endotracheal tube. The indicator changes from purple to yellow when exposed to a sufficient concentration of CO2 as a chemical reaction. The range of possible colors and the corresponding CO2 concentrations are listed in Table 8-1. Often it is difficult to evaluate the colors, which poses challenges in predicting the CO2 concentration in the expired gases. If the color changes to yellow during every exhaled breath, the CO2 concentration can be estimated to be more than 2%, or over 16 mm Hg. If the color change is not yellow, the expired CO2 level is less than desirable (Table 8-1).



Table 8-1


Exhaled CO2 Detection by Colorimetric Device
















Color CO2 Concentration
Purple < 0.5%
Tan 0.5%–2%
Yellow > 2%

Although colorimetric devices are used frequently by critical care providers and paramedics to confirm the presence of CO2 after endotracheal intubation, these devices have their limitations. For example, acidic contents in the stomach may turn the indicator yellow, giving a false-positive reading. Also, a patient who is intubated during cardiac arrest may not expel sufficient CO2 because of the lack of circulation, and the indicator would not change colors even with the endotracheal tube in the correct place.2


The current recommendation from the American Heart Association is to use waveform capnography instead of colorimetric devices even for emergency intubations, as well as during cardiopulmonary resuscitation.1


Waveform Analysis


Normal Capnogram


CO2 concentration (Figure 8-4) can be plotted against time (time capnogram), or expired volume (volume capnogram). A time capnogram is commonly used in clinical practice (see Figure 8-4). It is similar in shape in all healthy adults. Any abnormal shape other than the normal capnogram requires full investigation to determine if the alteration in the shape is caused by a physiologic or pathologic abnormality.


f08-04-9780323085120

Figure 8-4 A, Time capnogram showing segments, phases, and angles. Inspiratory segment is phase 0, and expiratory segment is divided into three phases: I, II, and III. Maximum value of carbon dioxide (CO2) at the end of the breath is designated as partial pressure of end-tidal carbon dioxide (PetCO2) and is about 35 to 36 mm Hg. It is lower than the partial pressure of arterial CO2 (PaCO2, 40 mm Hg) by about 5 mm Hg. B, Volume capnogram (PCO2 versus expired volume): Volume capnogram showing subdivisions of tidal volume. Area under the CO2 curve is effective alveolar ventilation. Area above the CO2 curve and the arterial PaCO2 line indicates physiologic dead space. A vertical line is drawn across phase II such that the two triangles p and q are equal in area. This divides physiologic dead space into anatomic and alveolar dead spaces. C, A time capnogram recorded during cesarean delivery general anesthesia showing phase IV. (From Kodali BS: Capnography outside of the operating rooms, Anesthesiology 118(1):192–201, 2013.)

To understand the components of a capnogram, a basic understanding of lung anatomy and physiology is essential. The trachea divides into the right and left main bronchi, the right side being shorter, wider, and more vertical. The implication is that a mainstem intubation is more likely to occur with the tube in the right main bronchus rather than in the left. Each mainstem bronchus divides into lobar bronchi, two on the left and three on the right. These bronchi further divide multiple times and eventually end in terminal bronchioles, each of which gives rise to respiratory bronchioles and alveoli. The conductive airways include everything from the trachea to the terminal bronchioles. The portion of lung past the terminal bronchioles, composed of the respiratory bronchioles and alveoli, participates in gas exchange. The alveolus is the most important structure participating in gas exchange, and the contribution of each alveolus to the capnogram depends on its ventilation and perfusion. Air flow and blood flow are not distributed evenly within the lung. The alveoli in the apex of the lung have high ventilation–perfusion (V/Q) ratios (more ventilation, less perfusion), and they contain relatively less CO2 after gas exchange. In contrast, the alveoli in the lower portions of the lung have low V/Q ratios (less ventilation, more perfusion), and they contain CO2-rich gas. During expiration, the CO2 from all alveoli moves through the respiratory tract to exit the trachea, or the endotracheal tube, to the measuring sensor (Figure 8-5).1,2,6


f08-05-9780323085120

Figure 8-5 Capnograms Following Sedation.

The height is decreased in C compared with A, and respiratory rate is decreased in D compared with B. (From Kodali, BS: Capnography outside the operating rooms, Anesthesiology 118(1):192-201, 2013.)

A time-based capnogram is a graphic depiction of the partial pressure of end-tidal CO2 (PetCO2) over time during a ventilatory cycle. It includes both inspiratory and expiratory segments, which allows for monitoring of the dynamics of inspiration and expiration.


 



Clinical Reasoning Pearl


Never assume the partial pressure of exhaled carbon dioxide PetCO2 values reflect partial pressure of arterial CO2 (PaCO2) values without waveform analysis. Any change in the waveform could be an indication of a change in the patient’s pulmonary status and warrants further evaluation. Loss of the waveform may signal loss of effective respirations.


The inspiratory segment consists of phase 0, where a sharp decline in CO2 occurs, compared with the end of expiration (see Fig 8-4, A). The expiratory segment is divided into three segments. It begins with phase I, which represents CO2-free gas being expelled from the conducting airways and breathing apparatus. Since this anatomic dead space does not participate in gas exchange, the CO2 concentration is initially zero at the beginning of expiration. Phase II consists of a rapid upstroke in CO2 concentration. In this segment, mixing of expired air from anatomic dead space and alveoli occurs to produce a positive deflection in the graph. Phase III is a plateau with a slightly positive slope that represents CO2-rich gas from alveoli. As stated above, inspiration yields no CO2 as long as no rebreathing occurs, and this accounts for the rapid drop in CO2 concentration from phase III to zero. A normal tracing always returns to baseline prior to the next expiratory cycle.


The alpha-angle is the angle between phase II and phase III, and it is affected by changes in the slope of phase III. The clinical implication of this angle is that it is an indirect indication of V/Q mismatch in the lungs. The beta-angle is the angle between phase III and the descending segment of the capnogram, or the beginning of phase 0. The beta-angle is usually 90 degrees.1 The maximum concentration of CO2 at the end of the breath is designated as ETCO2. If measured in millimeters of mercury, it is designated as PetCO2. It is usually about 35 to 36 mm Hg and lower than PaCO2 by about 4 to 5 mm Hg. The difference between PetCO2 and PaCO2 represents alveolar dead space (PaCO2 − PetCO2).


 



Clinical Reasoning Pearl


In a patient who is hemodynamically stable, the PetCO2 value can be used to estimate the arterial value of PaCO2, with the PetCO2 levels 1 to 5 mm Hg less than the PaCO2 levels.


Causes of increased PetCO2 values include situations in which CO2 production is increased, for example, hyperthermia, sepsis, and seizures, or situations in which alveolar ventilation is decreased, for example, respiratory depression.


Causes of decreased PetCO2 values include situations in which CO2 production is decreased, for example, hypothermia, cardiac arrest and pulmonary embolism, or situations in which alveolar ventilation is increased, for example, hyperventilation.


Capnograms in Nonintubated Patients


Distinct capnograms as described above may not be present in patients breathing spontaneously without an endotracheal tube. In the spontaneously breathing patient without an endotracheal tube, nasal sampling or sampling via a facemask may result in capnograms shown in Figure 8-6. These variations in shapes of capnograms are caused by the dilution of expiratory gases with atmospheric air or by supplemental oxygen administered via the mask. These capnograms can be considered normal variants in spontaneously breathing patients. Changes from baseline capnograms may reflect the effect of sedation on breathing.


f08-06a-9780323085120f08-06b-9780323085120f08-06c-9780323085120f08-06d-9780323085120f08-06e-9780323085120f08-06f-9780323085120f08-06g-9780323085120f08-06h-9780323085120f08-06i-9780323085120f08-06j-9780323085120f08-06k-9780323085120

Figure 8-6 Capnograms Shape and Clinical Significance.

A, Esophageal intubation. In esophageal intubation, the carbon dioxide (CO2) trace may have zero line or few blips of CO2 waveforms, depending on the concentration of CO2 in the esophagus. The presence of CO2 is caused either by swallowing of respiratory gases or by expiratory gases being pushed into the stomach during positive pressure ventilation via mask. However, with few ventilatory breaths, the CO2 concentration decreases to zero resulting in a flat line. B, Curare cleft. A sudden dip during phase III, or the expiratory plateau, indicates spontaneous respiratory effort during mechanical ventilation. It may occur with surgical manipulation or any other cause of brief external compression of the abdomen. It also indicates recovery from neuromuscular relaxant agents.2 C, Rebreathing. An elevation from the baseline is usually caused by an exhausted CO2 absorber, which results in rebreathing of CO2 during inspiration. As explained above in the analysis of the time capnogram, the CO2 concentration decreases to zero at the end of the alveolar plateau, reflecting the beginning of inspiration of CO2-free fresh gases. If rebreathing occurs, the downstroke will not reach the zero line and thus is elevated. During anesthesia, when controlled ventilation is used with closed circuit, elevation of baseline suggests inadequate CO2 absorption and exhausted soda lime. A defect in inspiratory and/or expiratory valve malfunction produces increases in the beta-angle, and/or elevation of the baseline. When a mechanical ventilator is used, an elevation from baseline indicates malfunction within the ventilator (faulty ventilator valves).6 D, Airway obstruction. An increase in the alpha-angle and an increase in the slope of phase III indicate airway obstruction. If severe enough, phase II may also be prolonged. The most common causes are kinking of the endotracheal tube, bronchospasm, asthma, or chronic obstructive pulmonary disease (COPD) or emphysema. Because of the airway obstruction, CO2 is expired at a slower rate, which is reflected by the more gradual increase in CO2 concentration.2 E, Normal variant of capnogram in pregnancy. The slope of phase III may be slightly increased in pregnant women undergoing general anesthesia. Although phase III in this figure looks similar to airway obstruction, phase II remains normal. This can be a normal physiologic variation in pregnancy.6 Hemodilution and increases in cardiac output produces better perfusion of alveoli thereby decreasing alveolar dead space. This brings the partial pressure of end-expiratory CO2 (PetCO2) closer to the partial pressure of arterial CO2 (PaCO2), and the difference between the two is less than in nonpregnant subjects. However, regional variation occurs in ventilation-to-perfusion (V/Q) ratios because of the pressure of the growing uterus toward the diaphragm. This results in alveoli with a lower V/Q ratio containing relatively more CO2 compared with the alveoli in the upper portions of the lung. The late emptying of the alveoli with higher CO2 concentration from the lower portions of the lung results in the increase slope of the phase III, or the alveolar plateau. F, Cardiogenic oscillations. The ripple effect at the end of phase III occurs during low-frequency mechanical ventilation. Each heartbeat creates rhythmic lung compression that moves the respiratory gases back and forth. The sensor sampling site detects these changes in CO2 to produce a ripple effect.6 G, Hypothermia. A decrease in end-tidal CO2 (PetCO2) occurs during periods of hypothermia, reduced metabolism, and low cardiac output. The height of the waveform is shorter, but other characteristics remain unchanged.6 H, Hypermetabolic state. A gradual rise in CO2 concentration may be caused by a hypermetabolic state. Rising PetCO2 is considered an ominous sign of malignant hyperthermia. A similar tracing may be seen with hypoventilation.6 I, Lung transplantation. A biphasic capnogram is seen in a patient that has had single lung transplantation. The initial peak in the waveform represents the transplanted, healthy lung, which has normal compliance and ventilation/perfusion (V/Q) ratio. The delayed, second peak of the waveform represents the native, diseased lung, which has poor compliance and significantly greater V/Q mismatch. The native lung takes longer to expel CO2 and mimics a capnogram from a patient with chronic obstructive pulmonary disease (COPD). The differing capnograms produced by each lung produces a dual peak capnogram. A similar capnogram may be seen when a partial disconnection or leak occurs within the apparatus.6 J, Inspiratory valve malfunction. Malfunction of the inspiratory valve leads to rebreathing of carbon dioxide (CO2) during inspiration. The waveform shows a slow decline in CO2 during inspiration and the tracing may not return to baseline prior to the next expiratory cycle, and the beta-angle is greater than the usual 90 degrees.6 K, Expiratory valve malfunction. Malfunction of the expiratory valve produces a shallow tracing that does not return to baseline with subsequent breaths. Phase II and phase 0 are abnormal because of mixing of inspiratory (CO2-free) and expiratory (CO2-containing) gases.6 (From Kodali, BS: Capnography outside the operating rooms, Anesthesiology 118(1):192–201, 2013.)

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Jun 3, 2017 | Posted by in Uncategorized | Comments Off on Capnography Monitoring

Full access? Get Clinical Tree

Get Clinical Tree app for offline access