Haemodynamic monitoring

Chapter 12 Haemodynamic monitoring



Haemodynamics is the study of blood flow. Haemodynamic monitoring therefore refers to the monitoring of blood flow through the cardiovascular system. In the intensive care unit (ICU), haemodynamic monitoring is used to detect cardiovascular insufficiency, differentiate contributing factors and guide therapy.


There has been debate over the risks and benefits of invasive haemodynamic monitoring in critically ill patients. Because there are scant data to show that invasive monitoring leads to improved survival, there has been a steady trend toward less invasive monitoring in the ICU. Regular clinical assessment remains an important component of haemodynamic monitoring, and any acquired physiological data must be interpreted in the clinical context. Importantly, monitoring must be combined with effective therapy in order to demonstrate an impact on outcome.1


As an aid to understanding haemodynamics, several circulatory models have been proposed, each with limitations. The standard model consists of a non-pulsatile pump and a hydraulic circuit with discrete sites of flow resistance. Heart rate and stroke volume determine cardiac output. The concepts of preload, contractility and afterload (Frank–Starling mechanism) are incorporated as determinants of stroke volume. In practice, quantification of these parameters is difficult, both clinically and in the laboratory. The model itself is simplistic, and does not allow for the pulsatile interaction of the cardiac pump with the elastance of the arterial tree. More complex models based on electrical circuits have been devised, but are still unable to quantify afterload and contractility precisely.


At the bedside the clinician must work with inexact surrogates of preload, contractility and afterload, measured or derived from arterial blood pressure (systemic or pulmonary), volume or pressure indices of cardiac filling, cardiac output and various markers of tissue well-being. In this chapter we focus on all these measurements except for those of tissue well-being, which are discussed elsewhere.



ARTERIAL BLOOD PRESSURE


The systemic pulse wave propagates from the aortic valve at 6–10 m/s. During its passage into the peripheral vasculature there is a progressive increase in systolic (SBP) and reduction in diastolic blood pressures (DBP), as standing and reflected waves become incorporated into the waveform, a process known as distal pulse amplification. Consequently, systemic arterial pressure measurements vary according to the site of measurement.


Mean arterial pressure (MAP) is arguably a more relevant index to monitor than either SBP or DBP for three reasons:





In the absence of sophisticated waveform analysis, arterial pressure is poorly correlated with cardiac output. Hence clinical estimation of cardiac output from arterial pressure and heart rate is unreliable until extreme hypotension occurs.2



NON-INVASIVE ARTERIAL BLOOD PRESSURE (NIBP) MEASUREMENT


In ICU, most standard NIBP instruments are automated intermittent oscillometric devices. Finger plethysmography and arterial tonometry can monitor both arterial pressure and waveforms continuously, but there are concerns regarding their accuracy.


Oscillometric NIBP is often used to check the reliability of invasive measurements, or can be used alone when beat-to-beat monitoring is not required. Contraindications to NIBP are relative, and influence the site of cuff placement. For instance, it is preferable to avoid extremities with severe peripheral vascular disease, venous cannulation, arteriovenous fistula or previous lymph node clearance (for example, radical mastectomy).


To make an oscillometric blood pressure measurement, a pneumatic cuff is inflated around a limb until all oscillations in cuff pressure are extinguished. The occluding pressure is then lowered stepwise, so that oscillations reappear over a discrete interval. Proprietary algorithms compute MAP, SBP and DBP from the alterations in oscillatory amplitude during deflation.


Oscillometry overestimates low pressures and underestimates high pressures, but for the normotensive range the 95% confidence limits are ± 15 mmHg (2 kPa). Dysrhythmias increase the error. Cuff width should be 40% of the mid-circumference of the limb. Narrower cuffs overestimate and wider cuffs underestimate blood pressures.


Complications are unusual. Repeated cuff inflations can cause skin ulceration, oedema and bruising, more so when the consciousness is impaired by illness and sedation. Ulnar nerve injury is also possible, especially with low cuff placement.



INVASIVE BLOOD PRESSURE MEASUREMENT3


Invasive blood pressure measurement is desirable in the presence of haemodynamic instability, end-organ disease requiring beat-to-beat blood pressure monitoring, during therapeutic manipulation of the cardiovascular system or if non-invasive methods fail. Cannulation of a systemic artery allows continuous monitoring of the arterial pressure waveform, heart rate and blood pressure, and also facilitates frequent arterial blood gas analysis. Relative contraindications include coagulopathy and vascular abnormality or disease.


The radial artery is the most common site for cannulation. Nothing larger than a 20-gauge cannula is advisable, and either a modified Seldinger technique or direct cannulation can be used. Once inserted, the cannula is usually infused with normal or heparinised saline at 3 ml/h, with a snap flush rate of 30–60 ml/h. Although it appears not to prolong patency, the use of heparinised saline (2–4 units/ml) might improve accuracy.4 It should be avoided if heparin-induced thrombocytopenia syndrome is suspected.


The axillary, brachial, femoral, posterior tibial and dorsalis pedis arteries can all be used without apparently raising the complication profile (Table 12.1). In severe circulatory compromise, gaining peripheral arterial access may be difficult and time-consuming. Rapid femoral cannulation by the Seldinger percutaneous technique usually remains feasible, with the added advantage that femoral arterial monitoring more accurately reflects aortic pressure in low-output states.


Table 12.1 Complications of arterial cannulation, with suggested preventive and treatment options61











































Complication Prevention Treatment
Vascular thrombosis (ranges from 7 to 30% following radial artery cannulation). Risk factors for digital ischaemia include: shock, sepsis, embolus of air or clot, hyperlipoproteinaemia, vasculitis, female sex, prothrombotic states, accidental intra-arterial injection of drugs Risks reduced by smaller catheter, larger artery, decreased duration of cannulation, avoiding traumatic insertion and multiple attempts. Allen’s test (including modifications such as Doppler, plethysmography and digital blood pressure) is probably unhelpful Remove cannula. Arterial thrombosis is usually self-limiting. Severe ischaemic damage estimated at < 0.01%. Anticoagulation and/or vascular surgery/intervention may be necessary
Distal embolisation Diligent catheter care and observation As for thrombosis
Proximal embolisation of clot or air (can result in stroke) Diligent catheter care and observation. Exclude air from pressurised system. Avoid axillary, subclavian or carotid access Tailored to sequelae
Vascular spasm Smaller catheter, larger artery. Avoid traumatic insertion and multiple attempts Remove cannula. Resite if necessary
Skin necrosis at catheter site Diligent catheter care and observation Surgical debridement and skin grafting may be necessary
Line disconnection and bleeding/exsanguination Minimise connections. Diligent catheter care and observation Control bleeding. Transfusion may be necessary
Accidental drug injection Clearly label arterial line near ports Leave cannula in situ to facilitate treatment if required. Depends on drug injected. May require papaverine or procaine, analgesia, sympathetic block of limb and anticoagulation
Infection – local or systemic Diligent catheter care and observation Remove cannula and send tip for culture. Resite if necessary. Immobilise and elevate affected upper limb. Start empiric antibiotics if sepsis or septic shock is present
Damage to nearby structures such as nerves, directly or due to haematoma (e.g. compartment syndrome or carpal tunnel syndrome). Arteriovenous fistula. Femoral approach can be associated with bowel damage Careful insertion technique. Seek assistance from experienced operator Tailored to sequelae. Haematomas can develop into pseudoaneurysms requiring surgery

Cannulae are usually removed or resited between days 5 and 7, or earlier if a complication is suspected. Distal perfusion should be checked at least 8-hourly, and the cannula removed if there is persistent blanching, coolness with sluggish capillary refill, loss of pulses or evidence of raised muscle compartment pressures. Complications associated with arterial cannulation are presented in Table 12.1.



PHYSICAL PROPERTIES OF CLINICAL PRESSURE MEASUREMENT SYSTEMS5,6


In the standard set-up, the arterial cannula is connected to a linearly responsive pressure transducer via fluid-filled non-compliant tubing < 1 m in length. Modern disposable transducers are precalibrated using electrical signals, and are not normally calibrated further against known pressures. The system is zeroed to the level of the phlebostatic axis, normally the mid-axillary line at the fourth intercostal space. Subsequent lowering of the transducer relative to this axis will cause pressure overestimation, while raising it will cause underestimation.


The natural resonant frequency of the system should ideally exceed 30 Hz (> 10 harmonics) for heart rates up to 180 beats/min (3 Hz) to prevent distortion of the biological signal by sine-wave system oscillations. Damping refers to any property of an oscillatory system that reduces the amplitude of oscillations. Factors that increase oscillation in the system, such as increased tubing length, diameter or compliance, cause underdamping. Overdamping tends to smooth the waveform, causing underestimation of SBP and overestimation of DBP, while MAP tends to be preserved. Contributing factors include clots, air bubbles and loose connections. Damping can be assessed clinically by the fast-flush test (Table 12.2).7


Table 12.2 Fast-flush test to assess dynamic response (damping) of the blood pressure-monitoring system













Make a paper record of transducer output
Snap the valve of the continuous flush system. This produces a square wave on the output trace
Repeat the process at least twice
Resonant frequency is the distance between successive peaks divided by the paper speed in millimetres per second
Damping is satisfactory (critical damping) if each snap test has two to three oscillation waves, with each wave one-third or less the size of the preceding wave

Upstream resistance or turbulence can result in a flow-dependent pressure reduction at the cannula site that differs from damping in that the MAP, SBP and DBP are all reduced. This is described as attenuation and is often observed in ‘positional’ arterial lines.



ADDITIONAL INFORMATION FROM THE ARTERIAL WAVEFORM


Measurement systems used in ICU do not precisely quantify the rate of arterial pressure rise and fall. Nevertheless, the wide pulse pressure of aortic regurgitation and the slowed upstroke of severe aortic stenosis may be detected. Pulsus paradoxus can readily be quantified in the spontaneously breathing patient. Systolic time intervals can provide an indication of ventricular contractility.


The systemic arterial pressure waveform can also assist in the prediction of fluid responsiveness (see Functional haemodynamic monitoring, below).



ESTIMATION OF STROKE VOLUME AND CARDIAC OUTPUT8


Estimation of stroke volume by analysis of the arterial pressure waveform, in particular various properties of the pulse pressure (pressure above diastolic) component, has been studied for many years. Pulse contour analysis calculates stroke volume from the area under the systolic portion of the arterial pressure waveform. After individual calibration, these monitors can determine stroke volume beat-to-beat. Cardiac output, derived by multiplying stroke volume by heart rate, appears to offer reliable continuous cardiac output monitoring, even during haemodynamic instability.9 However, accuracy in the presence of arrhythmia has not been validated.


Transpulmonary indicator dilution (discussed below) is the usual method of calibration. Pulse contour devices that use thermodilution or lithium ion calibration are commercially available.10 Comparisons with pulmonary artery catheter (PAC) cardiac output measurements have shown good agreement, with mean bias values ≤ 0.1 l/min and precision (SD of bias) of the order of 0.6 l/min.


Disadvantages of such techniques can include:







Recently developed proprietary algorithms continuously measure stroke volume and cardiac output without requiring calibration against another method.11 Additional clinical validation is awaited.



CENTRAL VENOUS CATHETERISATION (CVC)


CVC is indicated for monitoring of central venous pressure (CVP) and administration of certain drugs and parenteral nutrition. Modified catheters are also available which continuously monitor central venous oxygen saturation (ScvO2). Contraindications to CVC are relative and reflect potential complications of the procedure (Table 12.3). These should be considered in selecting the site for catheter insertion. Particular attention should be paid to the coagulation profile and factors which affect coagulation such as thrombolysis and activated protein C therapy.


Table 12.3 Complications of central venous catheterisation, with preventive and treatment options62







































Complication Prevention Treatment
Intravascular loss of guidewire Ensure guidewire is always secured. Avoid retracting guidewire into insertion needle. Seek assistance from experienced operator Interventional radiology or surgery may be required for retrieval
Air embolism Proper patient positioning, including Trendelenburg (head-down tilt) for jugular or subclavian insertion. Consider alternative insertion sites. Ensure connections are tight Left lateral Trendelenburg position. Administer 100% oxygen and ventilatory support. If catheter in place, tighten all connections and attempt to aspirate air. Basic/advanced life support if necessary
Dysrhythmias and conduction defects such as right bundle branch block Continuous electrocardiographic monitoring during insertion. Correct placement of catheter tip Withdraw or remove guidewire or catheter
Damage to nearby structures: Pneumothorax/haemothorax/chylothorax/hydrothorax, particularly with approaches to subclavian vein. Nerve injury such as phrenic, recurrent laryngeal, Horner’s syndrome. Arterial puncture – including injury to carotid, subclavian, aorta or pulmonary artery. Haematoma, pseudoaneurysm, arteriovenous fistula can result. Stroke may result from carotid artery injury. Tracheal injury. Femoral approach can be associated with bowel damage Identify risk factors such as previous surgery, skeletal deformity or scarring at site of insertion. Seek assistance from experienced operator. Select insertion site depending on impact of potential complications. Consider ultrasound guidance. Scheduled, routine replacement of catheter increases risk of mechanical complications Tailored to sequelae
Line disconnection and bleeding Minimise connections. Proper catheter care and observation Control bleeding. Transfusion may be necessary
Infection – local or systemic (including endocarditis) Aseptic insertion technique. Lower risk with subclavian compared to internal jugular or femoral insertion. Antimicrobial-impregnated catheters. Disinfect catheter hubs. Routine scheduled resiting appears not to reduce systemic infection rates. Remove catheter when no longer required Remove catheter; resite if necessary. Culture blood and catheter tip. Start empiric antibiotics if sepsis or septic shock is present
Superior vena caval erosion can result in haemothorax or cardiac tamponade Remove catheter when no longer required Early detection and surgical intervention
Thrombosis Remove catheter when no longer required. Subclavian insertion lower risk than internal jugular or femoral Anticoagulation, vascular or endovascular intervention may be required

Traditionally, puncture of the central vein is performed by passing a needle along the anticipated line of the vein with reference to surface anatomical landmarks (landmark method). Catheterisation is usually achieved via the subclavian, internal jugular or external jugular veins into the superior vena cava.12 The median cubital and basilic veins are used less commonly. Access to the subclavian vein is usually via the infraclavicular approach, but the supraclavicular approach is safe and reliable in experienced hands.


The use of ultrasound has been advocated to optimise the success rate of CVC insertion and minimise complications.13 Two-dimensional imaging can be used to localise the vein and define anatomy prior to placement of a CVC by standard landmark techniques. Ultrasound can also provide real-time, two-dimensional guidance during CVC insertion. Alternatively, audio-guided Doppler ultrasound can aid localisation of the vein and differentiate it from its companion artery during CVC insertion.


In 2002, the UK National Institute for Clinical Excellence issued guidelines recommending the use of ultrasound for elective catheterisation of the internal jugular vein, and consideration of ultrasound guidance in most clinical situations.14 These recommendations for ultrasound-guided internal jugular catheterisation are supported by meta-analysis.15 However, there are limited data supporting its use for subclavian and femoral veins.


The right tracheobronchial angle and carina are common radiological markers of insertion depth. In an attempt to reduce further the incidence of venous erosion (which is rare but potentially catastrophic), some practitioners place the catheter tip lower in superior vena cava or even in the upper right atrium, ensuring that the catheter is parallel to the long axis of the vein so that the tip does not abut the vein or heart wall end-on.16


Femoral venous catheterisation with radiological positioning of the tip close to the right atrium produces pressure measurements in good agreement with the subclavian CVP.17



CENTRAL VENOUS PRESSURE


Jugular venous pressure, CVP and right atrial pressure (RAP) are often used interchangeably. However, in situations associated with increased central venous resistance, such as central vein sclerosis, these pressures may not be the same.


The normal CVP in the spontaneously breathing supine patient is 0–5 mmHg, while 10 mmHg is generally accepted as the upper limit during mechanical ventilation. In health there is a good correlation between CVP and pulmonary artery occlusion pressures (PAOP), but this is lost in many types of critical illness such as pulmonary hypertension, pulmonary embolism, right ventricular infarction, left ventricular hypertrophy and myocardial ischaemia. The relationship between CVP and right ventricular end-diastolic volume (RVEDV: preload) is altered in critical illness by changes in right ventricular diastolic compliance and juxtacardiac pressures.18


Except at extreme values, static measures of CVP do not differentiate patients likely to respond to fluid therapy from non-responders. However, dynamic changes in CVP either in response to volume loading or related to respiration can assist in evaluating volume status.19 For instance, a steep increase in CVP following volume challenge suggests the heart is functioning on the plateau portion of the Frank–Starling curve. Severe hypotension with a low or normal CVP is unlikely to be due to acute pulmonary embolism, cardiac tamponade or tension pneumothorax.


Although fluid manometry is sufficient to measure the CVP, the frequency response is low and waveform analysis impossible (Table 12.4). Therefore, an electrical transducer system is normally used.


Table 12.4 Analysis of central venous pressure waveform























Condition Pressure changes Waveform changes
Tricuspid regurgitation Increased RA pressure Prominent v-wave, x descent obliterated, y descent steep
Right ventricular infarction RA and RV pressure elevated. RAP does not fall and may rise in inspiration Prominent x and y descents
Constrictive pericarditis RA, RV diastolic, PA diastolic and occlusion pressures elevated and equalised. RAP may rise in inspiration Prominent x and y descents
Pericardial tamponade RA, RV diastolic, PA diastolic and occlusion pressures elevated and equalised. RAP usually falls in inspiration y descent damped or absent

RA, right atrial; RV, right ventricular; RAP, right atrial pressure; PA, pulmonary artery.


In a trend toward less invasive monitoring, peripheral venous pressure and non-invasive estimates of CVP have demonstrated potential as alternatives to direct CVP measurement.20,21




PULMONARY ARTERY CATHETER


Right-heart catheterisation using a flow-directed balloon-tipped catheter was introduced by Swan and Ganz in 1970.23 The ability to monitor sophisticated haemodynamic and gas exchange variables at the bedside appealed to clinicians, and the PAC was rapidly accepted into routine critical care.


In 1996 a non-randomised cohort study of PAC use in American teaching hospitals appeared to show that, in any of nine major disease categories, PAC in the first 24 hours increased 30-day mortality (odds ratio 1.24, 95% confidence interval (CI) 1.03–1.49), mean length of stay and mean cost per hospital stay.24 A simultaneous editorial called for a moratorium on PAC use, and for a prospective multicentre trial.25


A subsequent Cochrane database systematic review of PAC monitoring in adult ICU patients incorporated data from two recent multicentre trials and 10 other studies.26 The pooled mortality odds ratio for studies of general ICU patients was 1.05 (95% CI 0.87–1.26) and for studies of high-risk surgery patients was 0.99 (95% CI 0.73–1.24). PAC monitoring had no impact on ICU or hospital length of stay (where reported). A recent multicentre trial incorporating protocolised haemodynamic management of patients with acute lung injury compared PAC-guided with CVC-guided therapy.27 There were no significant differences in 60-day mortality or organ function between groups. Overall, these data suggest that PAC monitoring in critically ill patients is not associated with increased mortality or with survival benefit.


On the other hand, recent data suggest a differential association between PAC use and mortality dependent upon severity of illness. Logistic regression analyses of data from a tertiary-care university teaching hospital28 and from the National Trauma Data Bank (American College of Surgeons)29 suggest that PAC use may be associated with survival benefit in the most severely ill/injured patients.


Overall, the PAC still finds application as a haemodynamic monitoring tool in the ICU and the operating room. However, its role is under increasing scrutiny, and continues to decline in favour of less invasive alternatives. It could be argued that future PAC use should be guided by studies to determine optimal management protocols and appropriate patient groups.26


Traditional indications for PAC monitoring have been to:





Contraindications for insertion reflect those for CVC. If known in advance, atypical cardiac or vascular anatomy, either congenital or secondary to trauma or surgery, should also be considered. Less invasive monitoring might provide the data that are sought. Also, PAC monitoring does not exclude simultaneous recourse to other monitoring techniques.



CATHETER INSERTION30


A 7.5–9 F 15-cm introducer sheath is first inserted by the Seldinger technique. The subclavian and internal jugular veins are most commonly used. Access is feasible with the 110-cm catheter via the median cubital, basilic and femoral veins. The external jugular veins can also be used, although difficulty may be encountered passing the introducer both into the vein and then subsequently below the clavicle into the subclavian vein.


Balloon volume is 1.5 ml. The balloon should be tested prior to insertion into the sheath. Deflation should always occur passively to minimise the risk of balloon damage. Before passage through the heart the balloon should be inflated to assist flow guidance and protect against myocardial injury and dysrhythmias. Inflation should not be forced, and should not alter the waveform prior to wedging. The right atrium is reached at 15–20 cm from the internal jugular vein, 10–15 cm from the subclavian vein, 30–40 cm from the femoral vein and 40 and 50 cm respectively from the right and left basilic veins. The right ventricle and pulmonary artery (PA) are then entered at additional 10-cm intervals, with a further 10 cm to PA occlusion.31 Waveforms seen as the catheter floats to the wedged position are shown in Figure 12.1.


< div class='tao-gold-member'>

Stay updated, free articles. Join our Telegram channel

Jul 7, 2016 | Posted by in CRITICAL CARE | Comments Off on Haemodynamic monitoring

Full access? Get Clinical Tree

Get Clinical Tree app for offline access