Hemodynamic instability (presyncope, angina, altered mentation, pulmonary edema) secondary to:
Bradydysrhythmias:
Sinus bradycardia
Sinus node dysfunction
Atrioventricular (AV) node conduction blocks (second and third degree)
Acute myocardial infarction with bifascicular block, alternating bundle branch block, new left bundle branch block (LBBB), escape rhythm <40 bpm
Malfunctioning permanent pacemaker
Electrolyte/metabolic disturbances (i.e., hyperkalemia) if medical therapies fail or are unavailable
Post–cardiac surgery (i.e., valve replacement)
Thoracic trauma (i.e., cardiac contusion)
Other (i.e., lyme carditis, endocarditis)
Tachydysrhythmias (overdrive pacing, typically transvenous only):
Supraventricular tachycardia (SVT)
Ventricular tachycardia (VT)
Transvenous Pacing—same as above, but also including:
Failure or nontolerance of transcutaneous pacing
Bridge to permanent pacemaker placement
High risk of progression to complete heart block
Overdrive pacing
CONTRAINDICATIONS
Absolute
Asymptomatic, stable rhythms (i.e., first-degree AV block)
Prosthetic tricuspid valve (transvenous only)
Relative
Severe hypothermia (may be physiologic bradycardia; can induce fibrillation)
Brady-asystolic arrest >20 minutes
Drug-induced dysrhythmias (although can be utilized as last resort if antidote fails)
TRANSCUTANEOUS PACING PROCEDURE
Landmarks
Pacer pads should preferentially be placed over the precordium anteriorly and in the interscapular paraspinous region posteriorly. Anterolateral pad placement is also acceptable.
Supplies
Cardiac monitor with pacemaker capabilities (if electrocardiogram [ECG] monitor is not part of the pacer unit, a separate monitor and adaptor will be required)
Adhesive pacing pads
ECG electrodes
Safety razor
Technique
Preparation
In conscious patients, reassurance and explanation of the procedure, including expectations for discomfort, are extremely important
Remove excess hair if time permits
Continuous cardiac and pulse oximetry monitoring, intravenous access, and bedside capability for resuscitation, including airway management, defibrillation, and arrhythmia treatment, should be at the bedside before initiation
Pacer/Electrode Placement
Pads are placed as shown in FIGURE 9.1
Anterior chest pacing pad (negative charge electrode) is placed over the point of maximal impulse
If access to the posterior chest wall is limited or difficult, posterior pacer pad may also be placed in cardiac apex/base position (identical to electrical cardioversion placement)
ECG electrodes should be placed in limb lead positions for monitoring
Pacing
Identify pacemaker mode on equipment and turn to “on”
Set heart rate to 70 bpm
Place and maintain one hand in pulse-check position (radial, femoral, or carotid) or observe noninvasive or invasive blood pressure response
In bradyasystole and unconscious patients, set current to 150 to 200 mA and lower in 10-mA decrements; set current at the lowest level that will consistently achieve mechanical capture
In stable and conscious patients, set current to 10 mA and raise in 10-mA increments until mechanical capture is achieved
Observe cardiac monitor for pacemaker spikes and electrical capture—“electrical capture” refers to narrow pacemaker spikes followed by typically wide ventricular complexes
Monitor constantly for “mechanical capture”—a palpable arterial pulse induced by pacemaker discharges or perfusing blood pressure by noninvasive or invasive blood pressure monitoring
Titrate sedation/analgesia/anxiolysis to allow for tolerance of ongoing pacing
Failure to achieve mechanical capture should prompt immediate preparation for transvenous pacer placement
Complications
Unrecognized ventricular fibrillation
Local discomfort
Cutaneous injury
SAFETY/QUALITY TIPS
Procedural
The most common causes of failure to capture in transcutaneous pacing are improper electrode placement or large patient size
For many patients, adequate amperage will not be possible without aggressive sedation/analgesia
Cognitive
Compared to transvenous pacing, transcutaneous pacing is painful and ineffective. Transcutaneous pacing should be thought of as a brief bridge to transvenous pacing or correction of the underlying disorder.
Electrical capture is not mechanical capture, and mechanical capture is what counts. Once electrical capture occurs, mechanical capture must be immediately verified using pulses, invasive arterial pressure monitoring, ultrasound, or (most conveniently) pulse oximetry.
Be careful not to mistake ventricular fibrillation or tachycardia for a paced rhythm