Radiation Safety and Use of Radiographic Contrast Agents in Pain Medicine




Overview of Radiation Safety


Pain practitioners have come to rely on fluoroscopy and, to a lesser but growing extent, on computed tomography (CT) to facilitate image-guided pain treatment techniques. Fluoroscopy and computed tomography employ ionizing radiation to produce the x-rays needed for imaging. Understanding the physics and biology underlying the biologic effects of ionizing radiation will help pain practitioners to minimize radiation exposure to their patients, other involved personnel, and themselves during image-guided injection. The basic elements of the fluoroscopy unit are illustrated in Figure 72.1 . X-rays emanate from an x-ray tube, typically positioned beneath the table and the patient to minimize radiation exposure. The x-rays pass through the table and the patient to strike the input phosphor of the image intensifier, where they are converted to visible light and, in turn, detected by an output phosphor that transfers the signal to a digital camera for visual display on a monitor or transfer to film. The size and shape of the x-ray beam can be adjusted after exiting the x-ray tube and before entering the patient, from side to side by an adjustable linear collimator or in a circular, concentric fashion by an iris collimator. The C-arm allows variation in the axis of the x-ray beam in numerous planes relative to the patient.




Figure 72.1


Diagram of the components of a typical fluoroscopy unit.

(Reproduced with permission from Rathmell JP. Atlas of Image-Guided Intervention in Regional Anesthesia and Pain Medicine. 2nd ed. Philadelphia: Lippincott Williams & Wilkins; 2012, Figure 2-1.)


The use of CT has become more commonplace, particularly among radiologists. With the advent of fluoroscopy units that can rotate around the patient and acquire images at numerous angles and then reconstruct the images in multiple planes, the distinction between CT-fluoroscopy and traditional CT has become blurred. These CT-fluoroscopy units yield data that can be reformatted in multiple planes and produce final images rivaling the quality of conventional CT. However, acquiring images in multiple planes with either CT-fluoroscopy of conventional CT requires significantly greater radiation exposure than conventional fluoroscopy. In most instances, the superior anatomic information provided does not warrant the routine use of these advanced imaging modalities (see the further discussion of radiation dose later in this chapter). Patients undergoing high-risk procedures where small variations in anatomy can alter the risk/benefit ratio of a given technique may benefit from CT-fluoroscopy or conventional CT guidance.




Basic Radiation Physics


Radiation is energy radiated or transmitted as rays, waves, or in the form of particles. X-rays are one portion of the spectrum of electromagnetic radiation. As x-rays pass through matter, they impart enough energy to dislodge electrons (ionizing radiation), yielding free radicals that can lead to harmful biologic effects. In radiography, it is the x-rays that penetrate the body without effect that emerge to strike an image intensifier where they are converted to visible light and can be displayed on a monitor or transferred to film, producing an image based on x-ray penetrability of various tissues.


Several factors and definitions are central to any basic understanding of radiation safety. The biologic effects of ionizing radiation are proportional to the time of exposure, whereas radiation exposure is inversely proportional to the square of the distance from the radiation source. Radiation exposure is expressed as roentgen (R) or coulomb per kilogram, whereas the energy absorbed from radiation is expressed as radiation absorbed dose (rad) or as gray (Gy). Because different types of radiation can have different biologic effects, units of exposure are converted from rad to radiation equivalent in man (rem) or Sievert (Sv). The units used to express radiation exposure are listed in Table 72.1 . For x-rays, 1 R ∼ 1 rad ∼ 1 rem.



Table 72.1

Units Used to Express Radiation Exposure and Dose
























Term Traditional Units SI Units Conversion
Exposure Roentgen (R) Coulomb/kg (C/kg) 1 R = 2.5 × 10 −4 C/kg
Radiation absorbed dose rad Gray (Gy) 100 rad = 1 Gy
Radiation equivalent in man rem Sievert (Sv) 100 rem = 1 Sv


The electrical input to the tube that generates the x-rays can be varied to produce x-rays that differ in number and energy. Increased current applied to the x-ray tube (expressed as milliamps or mA) produces more x-rays, and the more x-rays that strike the image intensifier, the darker the image. Lengthening the exposure time will also increase the number of x-rays reaching the image intensifier, thus variations in current and exposure time are expressed as mAs (mA × seconds). Increased voltage (expressed as kilovoltage peak or kVp) applied to the x-ray tube results in x-ray emission at higher energy levels (i.e., with greater ability to penetrate). In general, high kVp (75 to 125 kVp) and low mA (50 to 1,200 mA) are employed for fluoroscopy with short exposure times. This combination optimizes image quality while minimizing radiation exposure. High kVp/low mA combinations expose the patient to significantly less radiation than low kVp/high mA combinations. Modern fluoroscopy units typically employ automatic brightness control (ABC), which automatically adjusts kVp and mA to yield optimal brightness and contrast.


The x-rays generated during fluoroscopy are a form of ionizing radiation and have the potential to produce significant biologic effects. Small doses of ionizing radiation can produce molecular changes that take years to manifest in the form of cancerous transformation. Exposure to low doses of ionizing radiation is likely inconsequential because normal cellular mechanisms repair the damage. The International Committee on Radiation Safety Protection (ICRP) has produced estimates of the maximum permissible dose (MPD) of annual radiation to various organs ( Table 72.2 ). Exposure below these levels is unlikely to lead to any significant effects, but the ICRP recommends that workers should not receive more than 10% of the MPD.



Table 72.2

Annual Maximum Permissible Radiation Doses

























Area/Organ Annual Maximum Permissible Dose
Thyroid 0.5 mSv (50 rem)
Extremities 0.5 mSv (50 rem)
Lens of the eye 0.15 mSv (15 rem)
Gonads 0.5 mSv (50 rem)
Whole body 0.05 mSv (5 rem)
Pregnant women 0.005 mSv to fetus (0.5 rem)

Data from the National Council on Radiation Protection and Measurements (NCRP). Report No. 116. Limitation of exposure to ionizing radiation . Bethesda, MD: NCRP Publications; 1993.


Use of fluoroscopy for interventional procedures grew rapidly during the late 1980s, leading to increased concerns about radiation exposure. In 1994, the U.S. Food and Drug Administration (FDA) issued a public health advisory about serious radiation-related skin injuries resulting from some fluoroscopic procedures. Today’s equipment and techniques have reduced the risks of radiation exposure dramatically. Radiation exposure during a typical epidural steroid injection carried out with fluoroscopy and assuming the practitioner is at least 1 meter from the x-ray tube has been reported to be as low as 0.03 mR. In contrast, the typical entrance skin exposure during fluoroscopy ranges from 1 to 10 R per minute. A typical single chest radiograph leads to a skin entrance exposure of 15 mR. Thus, 1 minute of continuous fluoroscopy at 2 R per minute is equivalent to the exposure during 130 chest radiographs. Minimum target organ radiation doses that lead to pathologic effects are shown in Table 72.3 . Radiation dermatitis still occurs in fluoroscopists with unknown long-term consequences. Estimates of the relative radiation dose to the patient during use of fluoroscopy in comparison to other common diagnostic radiologic procedures are shown in Table 72.4 .



Table 72.3

Minimum Target Organ Radiation Doses to Produce Organ Pathologic Effects







































Organ Dose (rad) Dose (Gy) Results
Eye lens 200 2 Cataract formation
Skin 500 5 Erythema
700 7 Permanent alopecia
Whole body 200-700 2-7 Hematopoietic failure (4-6 weeks)
700-5000 7-50 Gastrointestinal failure (3-4 days)
5000-10,000 50-100 Cerebral edema (1-2 days)


Table 72.4

Comparative Radiation Doses for Common Diagnostic X-ray and Fluoroscopic Procedures




















































































X-ray—Chest 0.1 mSv (10 mrem)
X-ray—Mammography 0.42 mSv (42 mrem)
X-ray—Skull 0.1 mSv (10 mrem)
X-ray—Cervical spine 0.2 mSv (20 mrem)
X-ray—Lumbar spine 6 mSv (600 mrem)
X-ray—Upper GI 6 mSv (600 mrem)
X-ray—Abdomen (kidney/bladder) 7 mSv (700 mrem)
X-ray—Barium enema 8 mSv (800 mrem)
X-ray—Pelvis 0.6 mSv (60 mrem)
X-ray—Hip 0.7 mSv (70 mrem)
X-ray—Dental bitewing/image 0.005 mSv (0.5 mrem)
X-ray—Extremity (hand/foot) 0.005 mSv (0.5 mrem)
Fluoroscopy, intermittent (e.g., for lumbar transforaminal or facet injection) 0.007-0.03 mSv (0.7-3 mrem)
Fluoroscopy, high dose (three- to sixfold the radiation exposure of standard dose)
Fluoroscopy, continuous, pulsed mode 0.2-1 mSv/minute of exposure (20-100 mrem/minute of exposure)
Fluoroscopy, continuous 2-10 mSv/minute of exposure (200-1000 mrem/minute of exposure)
Fluoroscopy, continuous, high dose 10-20 mSv/minute of exposure (1000-2000 mrem/minute of exposure)
Fluoroscopy, continuous, digital subtraction 20-40 mSv/minute of exposure (2000-4000 mrem/minute of exposure)
Computed tomography—Head 2 mSv (200 mrem)
Computed tomography—Chest 7 mSv (700 mrem)
Computed tomography—Abdomen/pelvis 10 mSv (1000 mrem)
Computed tomography—Extremity 0.1 mSv (10 mrem)
Computed tomography—Angiography (heart) 20 mSv (2000 mrem)
Computed tomography—Angiography (head) 5 mSv (500 mrem)
Computed tomography—Spine 10 mSv (1000 mrem)
Computed tomography—Whole body 10 mSv (1000 mrem)
Computed tomography—Cardiac 20 mSv (2000 mrem)

Data adapted from American Nuclear Society. Radiation dose chart. Available at www.new.ans.org/pi/resources/dosechart . Accessed January 9, 2011.

Fluoroscopy exposure values are approximate and vary widely based on the region of the body examined and the body habitus of each patient. The values presented are extrapolated from data provided by Philips Medical Systems for the Pulsera 9-inch mobile C-arm and the following references: Wagner AL. Selective lumbar nerve root blocks with ct fluoroscopic guidance: technique, results, procedure time, and radiation dose. AJNR Am J Neuroradiol . 2004;25:1592-1594; and Mahesh M. The AAPM/RSNA Physics Tutorial for Residents. Fluoroscopy: patient radiation exposure issues. RadioGraphics . 2001;21:1033-1045.





Basic Radiation Physics


Radiation is energy radiated or transmitted as rays, waves, or in the form of particles. X-rays are one portion of the spectrum of electromagnetic radiation. As x-rays pass through matter, they impart enough energy to dislodge electrons (ionizing radiation), yielding free radicals that can lead to harmful biologic effects. In radiography, it is the x-rays that penetrate the body without effect that emerge to strike an image intensifier where they are converted to visible light and can be displayed on a monitor or transferred to film, producing an image based on x-ray penetrability of various tissues.


Several factors and definitions are central to any basic understanding of radiation safety. The biologic effects of ionizing radiation are proportional to the time of exposure, whereas radiation exposure is inversely proportional to the square of the distance from the radiation source. Radiation exposure is expressed as roentgen (R) or coulomb per kilogram, whereas the energy absorbed from radiation is expressed as radiation absorbed dose (rad) or as gray (Gy). Because different types of radiation can have different biologic effects, units of exposure are converted from rad to radiation equivalent in man (rem) or Sievert (Sv). The units used to express radiation exposure are listed in Table 72.1 . For x-rays, 1 R ∼ 1 rad ∼ 1 rem.



Table 72.1

Units Used to Express Radiation Exposure and Dose
























Term Traditional Units SI Units Conversion
Exposure Roentgen (R) Coulomb/kg (C/kg) 1 R = 2.5 × 10 −4 C/kg
Radiation absorbed dose rad Gray (Gy) 100 rad = 1 Gy
Radiation equivalent in man rem Sievert (Sv) 100 rem = 1 Sv


The electrical input to the tube that generates the x-rays can be varied to produce x-rays that differ in number and energy. Increased current applied to the x-ray tube (expressed as milliamps or mA) produces more x-rays, and the more x-rays that strike the image intensifier, the darker the image. Lengthening the exposure time will also increase the number of x-rays reaching the image intensifier, thus variations in current and exposure time are expressed as mAs (mA × seconds). Increased voltage (expressed as kilovoltage peak or kVp) applied to the x-ray tube results in x-ray emission at higher energy levels (i.e., with greater ability to penetrate). In general, high kVp (75 to 125 kVp) and low mA (50 to 1,200 mA) are employed for fluoroscopy with short exposure times. This combination optimizes image quality while minimizing radiation exposure. High kVp/low mA combinations expose the patient to significantly less radiation than low kVp/high mA combinations. Modern fluoroscopy units typically employ automatic brightness control (ABC), which automatically adjusts kVp and mA to yield optimal brightness and contrast.


The x-rays generated during fluoroscopy are a form of ionizing radiation and have the potential to produce significant biologic effects. Small doses of ionizing radiation can produce molecular changes that take years to manifest in the form of cancerous transformation. Exposure to low doses of ionizing radiation is likely inconsequential because normal cellular mechanisms repair the damage. The International Committee on Radiation Safety Protection (ICRP) has produced estimates of the maximum permissible dose (MPD) of annual radiation to various organs ( Table 72.2 ). Exposure below these levels is unlikely to lead to any significant effects, but the ICRP recommends that workers should not receive more than 10% of the MPD.



Table 72.2

Annual Maximum Permissible Radiation Doses

























Area/Organ Annual Maximum Permissible Dose
Thyroid 0.5 mSv (50 rem)
Extremities 0.5 mSv (50 rem)
Lens of the eye 0.15 mSv (15 rem)
Gonads 0.5 mSv (50 rem)
Whole body 0.05 mSv (5 rem)
Pregnant women 0.005 mSv to fetus (0.5 rem)

Data from the National Council on Radiation Protection and Measurements (NCRP). Report No. 116. Limitation of exposure to ionizing radiation . Bethesda, MD: NCRP Publications; 1993.


Use of fluoroscopy for interventional procedures grew rapidly during the late 1980s, leading to increased concerns about radiation exposure. In 1994, the U.S. Food and Drug Administration (FDA) issued a public health advisory about serious radiation-related skin injuries resulting from some fluoroscopic procedures. Today’s equipment and techniques have reduced the risks of radiation exposure dramatically. Radiation exposure during a typical epidural steroid injection carried out with fluoroscopy and assuming the practitioner is at least 1 meter from the x-ray tube has been reported to be as low as 0.03 mR. In contrast, the typical entrance skin exposure during fluoroscopy ranges from 1 to 10 R per minute. A typical single chest radiograph leads to a skin entrance exposure of 15 mR. Thus, 1 minute of continuous fluoroscopy at 2 R per minute is equivalent to the exposure during 130 chest radiographs. Minimum target organ radiation doses that lead to pathologic effects are shown in Table 72.3 . Radiation dermatitis still occurs in fluoroscopists with unknown long-term consequences. Estimates of the relative radiation dose to the patient during use of fluoroscopy in comparison to other common diagnostic radiologic procedures are shown in Table 72.4 .



Table 72.3

Minimum Target Organ Radiation Doses to Produce Organ Pathologic Effects







































Organ Dose (rad) Dose (Gy) Results
Eye lens 200 2 Cataract formation
Skin 500 5 Erythema
700 7 Permanent alopecia
Whole body 200-700 2-7 Hematopoietic failure (4-6 weeks)
700-5000 7-50 Gastrointestinal failure (3-4 days)
5000-10,000 50-100 Cerebral edema (1-2 days)


Table 72.4

Comparative Radiation Doses for Common Diagnostic X-ray and Fluoroscopic Procedures




















































































X-ray—Chest 0.1 mSv (10 mrem)
X-ray—Mammography 0.42 mSv (42 mrem)
X-ray—Skull 0.1 mSv (10 mrem)
X-ray—Cervical spine 0.2 mSv (20 mrem)
X-ray—Lumbar spine 6 mSv (600 mrem)
X-ray—Upper GI 6 mSv (600 mrem)
X-ray—Abdomen (kidney/bladder) 7 mSv (700 mrem)
X-ray—Barium enema 8 mSv (800 mrem)
X-ray—Pelvis 0.6 mSv (60 mrem)
X-ray—Hip 0.7 mSv (70 mrem)
X-ray—Dental bitewing/image 0.005 mSv (0.5 mrem)
X-ray—Extremity (hand/foot) 0.005 mSv (0.5 mrem)
Fluoroscopy, intermittent (e.g., for lumbar transforaminal or facet injection) 0.007-0.03 mSv (0.7-3 mrem)
Fluoroscopy, high dose (three- to sixfold the radiation exposure of standard dose)
Fluoroscopy, continuous, pulsed mode 0.2-1 mSv/minute of exposure (20-100 mrem/minute of exposure)
Fluoroscopy, continuous 2-10 mSv/minute of exposure (200-1000 mrem/minute of exposure)
Fluoroscopy, continuous, high dose 10-20 mSv/minute of exposure (1000-2000 mrem/minute of exposure)
Fluoroscopy, continuous, digital subtraction 20-40 mSv/minute of exposure (2000-4000 mrem/minute of exposure)
Computed tomography—Head 2 mSv (200 mrem)
Computed tomography—Chest 7 mSv (700 mrem)
Computed tomography—Abdomen/pelvis 10 mSv (1000 mrem)
Computed tomography—Extremity 0.1 mSv (10 mrem)
Computed tomography—Angiography (heart) 20 mSv (2000 mrem)
Computed tomography—Angiography (head) 5 mSv (500 mrem)
Computed tomography—Spine 10 mSv (1000 mrem)
Computed tomography—Whole body 10 mSv (1000 mrem)
Computed tomography—Cardiac 20 mSv (2000 mrem)