Monitoring Cardiovascular Function 1: ECG Monitoring



SETTING UP ECG MONITORING


The following measures should be observed when setting up ECG monitoring (Jevon 2009):



1. Explain the procedure to the patient.


2. Prepare the skin: ensure that the skin is dry, not greasy; if necessary use an alcohol swab and/or abrasive pad to clean (Resuscitation Council UK 2010). If necessary, shave off any excess hair. This will also make it less uncomfortable for the patient when the electrodes are removed.


3. Attach the electrodes following locally agreed guidelines. Switch the cardiac monitor on and select the required monitoring lead.


4. Ensure that the ECG trace is clear. Rectify any difficulties encountered (see below).


5. Ensure that alarms are set within safe parameters following locally agreed guidelines and according to the patient’s clinical condition.


6. Ensure that the cardiac monitor can clearly be seen.


7. Document in the patient’s notes that ECG monitoring has commenced.


Correct chest and limb electrode placement (Fig. 4.2) is crucial for obtaining accurate information from any monitoring lead (Thompson 2010). The electrode placement and monitoring lead selected for ECG monitoring will depend on the following factors:



  • Monitoring system (e.g. three- or five-wire monitoring system). If a five-wire system is being used a suggested ECG electrode placement is red (right shoulder), yellow (left shoulder), green (left lower thorax/hip region), black (right lower thorax/hip region) and white on the chest in the desired V position, usually V1 (Jevon 2009). If a three-wire system is being used a suggested ECG electrode placement is red (right shoulder), yellow (left shoulder) and green (left lower thorax/hip region), although this system is much less commonly used in clinical practice.
  • Goals of monitoring, e.g. if arrhythmia diagnosis is the goal.
  • Patient’s clinical situation, e.g. in cardiopulmonary resuscitation, the precordium should be left unobstructed in case defibrillation is required (Resuscitation Council UK 2010).


Fig. 4.2 Suggested ECG electrode placement using a five-wire monitoring system.


c04f002

EASI 12-Lead ECG Monitoring


The conventional 12-lead ECG using 10 electrodes attached to the limbs and chest is recognised as the current medical standard for the identification, analysis and confirmation of many cardiac abnormalities including cardiac arrhythmias and cardiac ischaemia/infarction.


If 12-lead ECG monitoring is undertaken on a continual basis, the benefits include:



  • Facilitating the accurate recognition of cardiac arrhythmias
  • Enabling the monitoring of the mid-precordial leads which is particularly important for the detection and management of ischaemia
  • Enabling the recording of transient ECG events of particular diagnostic or therapeutic importance
  • Enabling the differentiation between post-PTCA (percutaneous transluminal coronary angioplasty) ischaemia and occlusion.

Unfortunately the use of a conventional 12-lead ECG system using 10 electrodes for continuous cardiac monitoring is cumbersome and generally not practical in the clinical area. However, the EASI system, a new concept in 12-lead ECG monitoring, requires the use of only 5 electrodes (Fig. 4.3):



  • E electrode on the lower sternum at the level of the fifth intercostal space
  • A on the left midaxillary line on the same level as the E electrode
  • S electrode on the upper sternum
  • I on the right midaxillary line on the same level as the E electrode..


Fig. 4.3 EASI 12-lead ECG monitoring system.


(Reproduced by permission of Philips.)


c04f003

A fifth ground electrode can be placed anywhere.


The EASI system for 12-lead ECG monitoring using only 5 electrodes is less cumbersome and more practical than the standard 10-electrode system. It is therefore more comfortable for the patient. In addition it will not interfere with such procedures as cardiac auscultation, cardiopulmonary resuscitation (CPR), defibrillation and echocardiography.



ECG monitoring should complement not replace basic nursing observations of the patient. Treat the patient not the monitor.


POTENTIAL PROBLEMS WITH ECG MONITORING


There are potential problems that can occur when undertaking ECG monitoring and it is important to realise that standardisation is possible only if the ECG is performed in a standard way each time (Thompson 2010). It has been identified in the literature that there is a wide variation in the identification of correct lead placement (McCann et al. 2007) leading to inaccurate diagnosis and patients being exposed to potentially harmful therapeutic interventions (Rajaganeshan et al. 2008). Potential problems that may be encountered include the following.


The ‘Flat-Line’ Trace


Check the patient immediately. However, the most likely cause is mechanical. Check:



  • the condition of the patient
  • that the correct monitoring lead is selected (usually lead II)
  • that the ECG gain is set correctly
  • that the electrodes are ‘in date’ and the gel sponge is moist, not dry
  • that the electrodes are properly connected
  • that the leads are plugged into the monitor.

Poor-Quality ECG Trace


If the ECG trace quality is poor, check:



  • all the connections
  • the brightness display
  • that the electrodes are ‘in date’ and the gel sponge is moist, not dry (Jevon 2009)
  • that the electrodes are properly attached.

If there are still difficulties obtaining a clear ECG trace, wiping the skin with an alcohol wipe may help. As electrodes tend to dry out after about 3 days, they should be changed at least that often although every 24 hours may be optimum to maintain skin integrity (Perez 1996).


Interference and Artefacts


Poor electrode contact, patient movement and electrical inter­ference, e.g. from bedside infusion pumps, can cause a ‘fuzzy’ appearance on the ECG trace. Interference can be minimised by applying the electrodes over bone rather than muscle (Resuscitation Council UK 2006). The patient should also be reassured and kept warm.


Wandering Baseline


A wandering baseline (ECG trace going up and down) is almost always caused by poor electrode contact to the skin (Thompson 2010). If respiration is the cause and the problem is not transient, serial ECG recordings should be taken in both inspiratory and expiratory phases (Thompson 2010).


Small ECG Complexes


Sometimes the ECG complexes may be too small and unrecognisable. Possible causes include pericardial effusion, obesity and hypothyroidism. However, sometimes it can be caused by a technical problem. Check that the ECG gain is correctly set and that lead II is being monitored. Repositioning the electrodes or selecting another monitoring lead sometimes helps (Jevon 2009).


Incorrect Heart Rate Display


If the ECG complexes are too small, a false low heart rate may be displayed. Large T waves, muscle movement and interference can be mistaken for QRS complexes, resulting in a false high heart rate being displayed. The nurse must be alert to the possibility of inaccurate heart rate readings, which can in particular be caused by poor electrode contact and interference (Jevon 2009). To minimise the potential for inaccuracies, a reliable good-quality ECG trace should be obtained.


Skin Irritation


ECG electrodes can cause skin irritation and contact dermatitis has been well reported in the literature (Rühlemann et al. 2010). The electrode sites should be regularly examined and if the patient’s skin appears irritated select another electrode placement.


False Alarms


Frequent false alarms will undermine the rationale for setting alarms and can also cause undue anxiety for the patient. It is important to ensure that the alarms are correctly and sensibly set and that the ECG is accurate, reliable and of a high standard.



Best Practice – ECG Monitoring



Ensure adequate skin preparation


Use ECG electrodes that are in date, with moist gel sponge


Position ECG electrodes and select monitoring lead following locally agreed protocols


Set cardiac monitor alarms according to the patient’s clinical condition


Ensure that the ECG trace is accurate


Ensure that the cardiac monitor is visible


THE ECG AND ITS RELATIONSHIP TO CARDIAC CONTRACTION


The ECG functions in four stages as follows (Fig. 4.4):



1. The sinus node fires and the electrical impulse spreads across the atria. This results in atrial contraction (P wave).


2. On arriving at the atrioventricular (AV) junction the impulse is delayed, allowing the atria time to contract fully and eject blood into the ventricles. This brief period of absent electrical activity is represented on the ECG by a straight (isoelectric) line between the end of the P wave and the beginning of the QRS complex. The P–R interval represents atrial depolarisation and the impulse delay in the AV junction before ventricular depolarisation.


3. The impulse is then conducted down to the ventricles through the bundle of His, right and left bundle branches and Purkinje fibres, causing ventricular depolarisation and contraction (QRS complex).


4. The ventricles then repolarise (T wave).

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Sep 1, 2016 | Posted by in EMERGENCY MEDICINE | Comments Off on Monitoring Cardiovascular Function 1: ECG Monitoring

Full access? Get Clinical Tree

Get Clinical Tree app for offline access