Mechanical Ventilation



Mechanical Ventilation


Mark J. Heulitt

Courtney Ranallo

Gerhard K. Wolf

John H. Arnold





In simple terms, the lung-ventilator unit can be thought of as a tube with a balloon network on the end, with the tube representing the ventilator tubing, endotracheal tube (ETT), and airways and the balloon network representing the alveoli. During mechanical ventilation, the forces generated are a combination image of effort provided by the patient’s muscles during spontaneous breathing attempts and support derived from the ventilator. The movement of gas is determined by forces, displacements, and the rate of change of component displacements, which are distensible. To generate a volume displacement, the total forces have to overcome both the elastic and resistive elements of the lung and airway/chest wall. To generate gas flow, the total forces must overcome the resistive forces of the airway, the ventilator tubing, and the ETT to allow gas to flow into or out of the lung, depending on driving pressure gradients.

This discussion of mechanical ventilation will first focus on an understanding of how the patient’s physiologic changes occur to generate these forces, both by intrinsic patient effort and from forces generated by ventilator output. How the ventilator controls the variables of flow, volume, and pressure to image generate these forces and how these controllers interface with the patient will then be examined. Finally, the indications, settings, and modes of ventilatory support available to clinicians who care for infants and children will be discussed.


PHYSIOLOGY OF MECHANICAL VENTILATION

In physiology, force is measured as pressure [pressure = force / area], displacement is measured as volume (volume = area × displacement), and the relevant rate of change is measured as flow (e.g., average flow = Δvolume / Δtime; istantaneous flow = dv / dt, where d represents the derivative of volume with respect to time). The key components in positive-pressure mechanical ventilation are the pressure necessary to cause a flow of gas to enter the airway and increase the volume of gas in the lungs. The volume of gas (ΔV) going to any lung unit (the balloons in our simplified example) and the gas flow (V·) may be related to the applied pressure (ΔP) by

ΔP = ΔV/C + [V with dot above] · R + κ

where R is the airway resistance, C is the lung compliance, and κ is a constant that defines end-expiratory pressure.

The above equation is known as the equation of motion for the respiratory system. The sum of the muscle pressures and the ventilator pressure is the applied pressure to the respiratory system. The muscle pressure represents the pressure generated by the patient to expand the thoracic cage and lungs. Unfortunately, this force is not able to be measured directly. In contrast, the ventilator pressure is the transrespiratory pressure (from the mouth to the alveolar space) generated by the ventilator during inspiration. Combinations of these pressures are generated when a patient is breathing on a positive-pressure ventilator. For example, when respiratory muscles are at complete rest, the muscle pressure is zero; thus, the ventilator must generate all of the pressure necessary to deliver the tidal volume (VT) and inspiratory flow. The reverse is also true—when the patient is making some respiratory effort and generating muscle pressure, the degree of support that must be supplied by the ventilator is less. Multiple modes of ventilator support can be used to assist the patient in this circumstance. With the basic principle behind the equation of motion defined, it must be applied to the forces that must be overcome to generate gas flow. The total pressure applied to the respiratory system (PRS) of a ventilated patient is the sum of the pressure generated by the ventilator (measured at the airway, PAO) and the pressure developed by the respiratory muscles (PMUS).


where VT is the tidal volume, CR the respiratory system compliance, (VT/CR) the elastic force, [V with dot above] the flow, R the airway resistance, ([V with dot above] × R) the resistive force, and PEEPtotal the total alveolar end-expiratory pressure.

PAO and [V with dot above] can be measured by the pressure and flow transducers in the ventilator. Volume is derived mathematically from the integration of the flow waveform.

To generate a volume displacement, the total forces have to overcome elastic and resistive elements of the lung and airway/chest wall, represented by VT/CR and [V with dot above] × R, respectively. VT/CR (elastic force) depends on both the volume insufflated in excess of resting volume and on respiratory system compliance. To generate a flow of gas, the total forces must overcome the resistive forces of the airway and the ETT against the driving pressure gradients. Given the physics principle that any action has an opposing reaction, at any moment during inspiration, a balance is maintained of forces attempting to expand the lung and the chest wall and those opposing lung
and chest wall expansion. We commonly measure the result of these forces as the airway pressure (PAWO). Opposing pressures to lung and chest wall expansion are the sum of elastic recoil pressure (Pelastic), flow resistive pressure (Presistive), and inertance image pressure (Pinertance) within the respiratory system. Thus,

PAWO = Pelastic + Presistive + Pinertance

Inertial forces relate to the energy required for initiation of gas movement and are usually negligible during conventional ventilation; thus, this component is commonly ignored. For conventional ventilation, the forces exemplified in the equation of motion can be expressed as follows:

PAWO = Pelastic + Presistive

If the elastic forces are the product of elastance and volume (Pelastic = E × V), and resistive forces are the product of flow and resistance (Resistive = [V with dot above] × R), and the formula can be written as:

PAWO = (Elastance × Volume) + (Resistance × Flow)

Elastance is the inverse of compliance. If compliance is substituted for elastance in the equation, the equation of motion results:


It is important to note that the quotient of volume displacement over compliance of the respiratory system represents the pressure necessary to overcome the elastic forces above the resting lung volume (known as the functional residual capacity, FRC). The FRC represents the quantity of air remaining in the lungs at the end of a spontaneous expiration. Pressure, volume, and flow are all measured relative to their baseline values. Thus, for a patient on a ventilator, the pressure necessary to cause inspiration is measured as the change in airway pressure above positive end-expiratory pressure (PEEP), representing the change from baseline pressure to peak inspiratory pressure. For example, in a patient breathing spontaneously on continuous positive-airway pressure (CPAP), the ventilator pressure is zero; thus, the patient must utilize respiratory muscles to generate all of the work of breathing (WOB) and the force necessary to expand the lungs and chest, to enable forward gas flow into the alveoli. The same principle can be applied to the volume generated during inspiration (the VT), which is the change in volume in the lung during inspiration above FRC. The pressure necessary to overcome the resistive forces of the respiratory tract is the product of the maximum airway resistance (Rmax) and the inspiratory flow. Flow is measured relative to its end-expiratory value and is usually zero at the beginning of an inspiratory effort, unless an intrinsic PEEP (PEEPi) is present. In this circumstance, flow may still be occurring within the lung as alveoli attempt to achieve their baseline state—that is, due to time-constant differences, overfilled alveoli may be emptying into underfilled alveoli to help both to achieve their “best” resting volume. This effect is known as pendelluft. When PEEPi is present, it will take more “effort” from the patient and the ventilator to generate enough flow to move gas into the lung.

At this point in the discussion of mechanical ventilators, those factors that are directly clinician-controlled must be differentiated from those that occur indirectly. For example, pressure, volume, and flow are directly controlled image variables, while other important factors such as resistance and compliance are dependent upon the resistive and elastic properties of the respiratory system and cannot be directly image controlled.


VENTILATOR CONTROLLERS

Each ventilator is essentially a controller of pressure, volume, or flow in the equation of motion. The manner in which each variable is controlled, described as the mode of ventilation, image determines how the ventilator delivers the mechanical breath. In the equation of motion, the form of any of the variables (pressure, volume, and flow are expressed as functions of time) can be predetermined. This principle serves as the theoretical basis for classifying ventilators as pressure, volume, or flow controllers (Fig. 38.1). The necessary and sufficient criteria for determining which variable is controlled are listed in Table 38.1. It is important to recognize that any ventilator can only directly control one variable—pressure, volume, or flow—at a time. Thus, a ventilator is simply a technology that controls the airway pressure waveform, the inspired volume waveform, or the inspiratory flow waveform, and pressure, volume, and flow are referred to in this context as control variables.

Most clinicians think of ventilators in terms of modes of ventilation. However, the mode of ventilation is meant to be a description of the way in which a mechanical breath is delivered. The determinants of how a mechanical breath is delivered are summarized not only in the control variables, but also in the phase and conditional variables (Fig. 38.1). Again, control variables are the independent variables that are either pressure, volume, or flow. Conditional variables are determinants of a response to a preset threshold, which are both clinician set and influenced by dependent and independent variables. Phase variables are those that are used to start, sustain, and end the phase. During inspiration, the phase variables include the trigger variable (determines the start of inspiration), limit variable (determines what sustains inspiration), and cycle variable (determines the end of inspiration).






FIGURE 38.1. Flowchart to emphasize that each breath may have a different set of control and phase variables, depending upon the mode of ventilation. (From Chatburn RL. Classification of mechanical ventilators. Respir Care 1992;37:1009-25, with permission.)









TABLE 38.1 VENTILATOR CONTROLLERS







































































FLOW CONTROLLER (CONSTANT-FLOW CONTROLLER)


PRESSURE CONTROLLER (CONSTANT-PRESSURE CONTROLLER)


VOLUME CONTROLLER (VARIABLE FLOW CONTROLLER)


Modes


Modes


Modes


VC, SIMV-VC


PC, SIMV-PC, PRVC


VC


Equation


Equation


Equation


image


image


Flow = Pressure × Compliance



Pressure = Resistance × Flow


Independent variables


Independent variables


Independent variables


Flow


Pressure


Volume


Dependent variables


Dependent variables


Dependent variables


Pressure


Volume


Pressure



Flow


Limiting variables


Limiting variables


Limiting variables


Volume


Pressure


Volume


Trigger variables


Trigger variables


Trigger variables


Time


Time


Time


Pressure


Pressure


Pressure


Flow


Flow


Flow


VC, volume control; SIMV-VC, synchronized, intermittent mandatory ventilation-volume control; PC, pressure control; SIMV-PC, synchronized, intermittent mandatory ventilation-pressure control; PRVC, pressure-regulated volume control.



Control Variables

Control variables relate to the elastic and resistive forces that must be overcome to allow gas delivery to the patient. An initial discussion of the elastic components of the equation of motion as it relates to pressure will help to simplify the explanation. It is known that compliance relates to the change in volume in the lung as a result of a change in pressure. Thus, pressure is related to volume and to the patient’s compliance.


If the clinician sets pressure as the control variable, volume varies directly with the compliance of the respiratory system. Thus, pressure is the independent variable set by the clinician, and volume is the dependent variable determined by the level of pressure. When the pressure pattern is preset by the clinician, the ventilator operates as a pressure controller. The volume becomes a function of compliance, so that a decrease in compliance allows less volume to be delivered for the same pressure. During expiration, the elastic and resistive elements of the respiratory system are passive, and expiratory waveforms are not directly affected by the modes of ventilation or the controller. However, as the respiratory cycle is a set period of time, any change in the inspiratory time can influence expiratory time and, to a certain extent, the expiratory profile.

For the resistive components of the equation of motion,

Pressure = Resistance × Flow

When a ventilator operates as a constant-pressure controller—for example, in pressure-control (PC) mode, pressure-regulated volume-control (PRVC) mode, and synchronized, intermittent mandatory ventilation-pressure control (SIMV-PC) mode—pressure is an independent, or controlled, variable (Table 38.1). The set pressure will be delivered and maintained constant throughout inspiration, independent of what resistive or elastic forces of the respiratory system might be. Even though pressure is constant, the delivered VT will vary as a function of compliance and resistance, and the flow will also vary exponentially with time.

A waveform from a ventilator operating as a pressure controller is displayed in Figure 38.2. Under this condition, volume and flow become the dependent variables, and their patterns will depend upon compliance and resistance. When a pressure pattern is preset (constant in a PC mode), flow-time and volume-time waveforms vary exponentially with time and are a function of compliance and resistance.

It is now clear that flow and resistance are associated only with the resistive components of the equation of motion. The elastic component refers to volume and compliance. Considering the resistive components of the equation of motion,


Thus, if the clinician sets flow as a function of time, pressure then varies with resistance. Flow is the independent variable and pressure is the dependent variable. When a flow pattern is preset, the ventilator operates as a flow controller; pressure is a function of resistance, and the inspiratory pressure-time waveform varies linearly with time. Volume increases linearly with time, although it does not have a direct relation to flow. Based on the equation of motion, volume does have an indirect relationship to flow, as volume is the integral of flow and flow is the derivative of volume. Again, expiration is passive, and the expiratory profile is not directly affected by mode of ventilation, but rather by compliance and resistance, even though the set inspiratory time can influence the expiratory time and, to a certain extent, the expiratory profile.

When a ventilator operates as a constant-flow controller (volume-controlled, VC, and SIMV-VC modes), flow is the independent variable. Regardless of the resistive or elastic forces of the respiratory system, the set flow will be delivered
and maintained constant throughout inspiration. Pressure and VT will vary with time, depending on the compliance and resistance.






FIGURE 38.2. Volume-time wave form from a constant-pressure mode of ventilation. A: Increased resistance. With inspiration, an abnormal increase in tidal volume and decrease in inspired tidal volume occur, as compared to tracing B (normal resistance). Expiration has an abnormal linear decay to baseline. B: Normal resistance. During inspiration, a normal exponential increase in tidal volume occurs. Expiration has a normal exponential decay to baseline.

Waveforms from a ventilator operating as a flow controller are illustrated in Figure 38.3. Flow is the independent variable (controlled variable); pressure and volume are dependent variables. When a flow pattern is preset (constant in this case), pressure and volume are the dependent variables; they vary linearly with time and are functions of compliance and resistance. The modern ventilator can operate as a flow controller or as a pressure controller. As a flow controller, the most common pattern is constant flow, also referred to as a square-wave flow pattern. In this mode, the flow increases to a set level that is maintained for the duration of the inspiratory time. The pressure and volume that the patient receives are a function of compliance and resistance. When a ventilator is set as a pressure controller, the observed pressure pattern is constant and results in a square-wave pressure pattern.

From the equation of motion, it can be stipulated that, with the ventilator operating as a constant-flow controller, the pressure and volume are linear functions of time. The various ventilators are able to deliver various flow patterns. To have flow patterns different from the most common types, which are constant and exponentially decelerating, the ventilator must be controlled by a microprocessor, which performs a series of sequential adjustments dictated by an algorithm to produce various flow patterns, including decelerating ramp, ascending ramp, and sinusoidal. These flow patterns are used in various volume-cycled modes. Again, volume-cycled ventilation is still controlled by flow, and the independent variable set is flow. It is important to note that the decelerating rate of flow is controlled by an algorithm that does not reflect the elastic and resistive elements of the respiratory system. Abnormalities in these elements may result in flow-starvation asynchrony, as the flow from the ventilator may be inadequate to meet patient needs. Not all ventilators can provide novel flow patterns.






FIGURE 38.3. Pressure-time waveform from a constant-flow mode illustrating resistive and elastic elements of the respiratory system.

Describing the flow and pressure patterns observed in different modes of ventilation is often confusing jargon. To expand further, in PC ventilation, the pressure pattern is “square,” but flow increases rapidly at the beginning of the inspiratory phase
to generate the set pressure limit, then decays exponentially over the inspiratory time. This flow pattern is described as decelerating flow. It is said that the major difference between volume and pressure ventilation is based on the square-wave and the decelerating flow patterns observed. In PC mode, the initial “snap” of high flow to reach the set pressure limit has been thought to be potentially beneficial in opening stiff alveoli in conditions such as acute respiratory distress syndrome (ARDS) or surfactant deficiency. It has been proposed that a decelerating flow favors better gas exchange and improves distribution of ventilation among lung units with heterogeneous time constants. For this reason, clinicians often choose PC ventilation in patients with poor compliance, although the true benefit of PC versus other modes has not been well established in animal or clinical studies.

Volume is most closely associated with the elastic component of the equation of motion. The resistive component of the equation is most dependent on resistance and flow. The image elastic components can be rearranged to display how volume is determined:

Volume = Pressure × Compliance

If the clinician sets volume as a function of time, pressure then varies with compliance. Volume is the independent variable and pressure is the dependent variable.

Theoretically, when a ventilator sets a volume pattern, it operates as a volume controller. However, to be a true volume controller, the ventilator must measure volume directly in order to set the volume pattern. Most ventilators cannot directly measure volume; rather, they calculate volume delivered from flow that occurs over a period of time. Most ventilators use volume as a limiting variable, meaning that inspiration stops when the preselected volume is reached. When inspiration stops at the preset volume value, the ventilator is referred to as volume cycled, but it is really acting as a flow controller. That is, the ventilator is set to deliver a set VT at a certain number of breaths per minute and a specified inspiratory time, which determines the patient’s minute ventilation. The ventilator gives a certain flow to meet the set requirements.

Understanding the delivery of a positive-pressure breath to a patient requires an understanding of the relationship of gas delivery to overcoming the resistive and elastic elements of the patient and the ventilator system.


Volume Measurement

The goals of modern mechanical ventilation in infants and children have focused on preventing overdistension of alveoli by limiting VT, thus reducing volutrauma (1,2,3). Exact knowledge of both inspired and expired gas volumes with a sufficient image level of precision is essential to optimize ventilator settings using lung-protective techniques. During the inflation phase of mechanical ventilation, pressure rises within the ventilator circuit, causing elongation and distension of the tubing and compression of the gas within the circuit. The volume of the gas, which becomes compressed within the ventilator tubing and never reaches the patient, is termed the compressible volume of the circuit. While in most circumstances, this volume is standard within different sizes of circuits, variation can occur. Knowing the compressible volume within the ventilator circuit is important in determining the actual VT being delivered to the patient’s lungs. When compression volume is accounted for in determining patient VT, the resultant volume is termed image the effective tidal volume (eVT), which means that this is the VT that reaches the patient’s lungs. eVT can be calculated thus:

eVT = (VTE ) – [Circuit compensation × (PImaxPEEP)]

where VTE is the expired tidal volume and PImax is the maximal inspiratory pressure.

The optimal site for monitoring volumes in infants and children is unclear. The inability to accurately measure VT in a image conventional ventilator is caused by several factors, including (a) difficulty of compensating for volume loss in the ventilator circuit or in the humidifier and (b) changes in temperature, humidification, and secretions, which may also influence the amount of gas that gets delivered from the ventilator to the actual patient. Air leaks around the ETT itself, especially in small patients with uncuffed tubes, are another source of volume measurement error. Measuring VT at the proximal airway eliminates most circuit compliance and other dead-space factors. Therefore, it has been recommended that the proximal airway be the only site at which to obtain accurate volume measurements in infants and children (4). To measure volumes at the proximal airway, a pneumotachograph must be positioned at the patient’s airway opening or at the ETT. Unfortunately, this technique has disadvantages that are especially apparent in infants and children. A pneumotachograph placed at the proximal ETT opening creates dead space of its own, which can be detrimental in infants who already have small VTs (5). In addition, the proximally placed pneumotachograph may impair admittance to the ETT and airways and make suctioning more difficult. Secretions can also result in contamination of the pneumotachograph and distort observed measurements. Finally, the weight of the pneumotachograph at the proximal end of the ETT increases its overall weight and may result in an increased risk of extubation.

Given the importance of knowing the true VT delivered to a patient, it is essential to determine which is the most accurate, safe, and efficient site to monitor. It is also important to understand the reliability of conventional displays of VT generated by most ventilators. Three clinical studies have compared the ventilator-displayed VT (measured at the exhalation valve for exhaled VT displays) with the VT measured at the ETT. One group studied 98 ventilated infants and children using the Servo 300TM ventilator; 70 patients were ventilated with an infant circuit (compliance: 0.61 mL/cm H2O), and 28 patients were ventilated with a pediatric circuit (compliance: 1.0 mL/cm H2O) (6). In the patients with the infant circuit, poor correlation was noted between the expiratory VT measured with the pneumotachometer and the ventilator-displayed VT (R2 = 0.54). Poor correlation (R2 = 0.58) was also noted between the expiratory VT measured with the pneumotachometer and the calculated effective VT. In pediatric circuits, greater correlation occurred between the pneumotachometer-measured expiratory VT, the ventilator-displayed VT, and the calculated effective VT (R2 = 0.84 and 0.85, respectively).

A second group studied 54 ventilated infants and children using the Servo 300TM ventilator with a pediatric circuit (compliance: 1.35 mL/cm H2O) and adult circuit (compliance: 2.4 mL/cm H2O) (7). The VT measured at the ETT was significantly less than the ventilator-measured tidal VT and varied between 2% and 91%, with a mean (standard deviation) error of 32% (20%) in 40 children with the pediatric circuit and 18% (6%) in 16 children with the adult circuit. The VTs displayed by the ventilator were also significantly different from the calculated effective VT (-63.3% to +29.1%), with substantial underestimation of VT from the VT displayed or the effective VT in many patients with small VTs. The third group studied 30 ventilated infants, 1-23 months old, on Servo 300. ventilators with neonatal (compliance: 0.63 mL/cm H2O) and pediatric (compliance: 1.13 mL/cm H2O) circuits (8). This study demonstrated that the expiratory VT measured at the ETT was less than the expiratory VT displayed by the ventilator. When the ventilator was in the VC mode, the median difference was -36% (range: -5% to -2%), while the median was -35% (range: -6% to -60%) in the PC mode. The calculated
effective VT was not statistically different from the ventilator-measured VT in PC mode; however, individual differences were large (range: -26% to +52%). The calculated effective VT was less than the VT measured at the ETT in the VC mode, with large individual differences (median: -23%, range: -48% to +21%). All of these clinical studies have recommended that the VT should be measured at the ETT in infants and small children. As stated previously, the effective VT can be calculated as the ventilator-measured expired VT. However, this method fails to take into account volume lost internally in the ventilator. Manufacturers have attempted to compensate for these volume losses by measuring compression volume loss in the system. The compliance factor can be calculated as


where Kc is the compliance factor, dv is the total integrated volume, and P0 and P1 are the start and target pressures.

Data demonstrate good agreement in volume measured at the ventilator when compensation adjustment is on, as compared to volume measured at the proximal airway (9). The use of circuit-compliance compensation improved the agreement between the volume measured by the ventilator in an animal trial; pediatric pigs had improved agreement between the two volume methods attributable to circuit-compliance compensation (with circuit-compliance compensation “on,” 0.97; with circuit-compliance compensation “off,” 0.88; p = 0.027). It is essential for the clinician to understand the accuracy of the delivery of VT to their patients, especially in patients with small volumes, such as infants.

In a clinical study of 68 ventilated pediatric patients aged between 2 days and 18 years, the principal observation was that, when compression volume was compensated for by a computer algorithm, a negative bias occurred. Displayed volume was lower than that set by the clinician, but agreement was good when using compensation, with a concordance correlation between 0.90 and 0.98. Accuracy, expressed as percent difference, improved only in older patients (10). A limitation of this study, as well as previous studies that utilized an airway sensor or pneumotachograph as the reference value of volume, is the short time frame in which volumes were measured. A potential limitation of the airway sensor is the increased opportunity for the sensor to be contaminated by airway secretions and foreign substances, as compared to a pneumotachograph located within the ventilator. The potential for contamination, and thus degradation of the accuracy of the signal, of an airway sensor increases with increased length of time that the sensor is in place. In all of these studies, the airway sensor was only in place for a limited period. Caution must be used in extrapolating measurements from previously cited studies when the airway sensor is not left in place over a prolonged period.

When examining volume inaccuracies, the first source of error would be loss of volume within the ventilator circuit resulting from compression volume of the circuit. In the inflation phase, pressure rises within the ventilator circuit causing elongation and distention of the tubing and compression of the gas within the circuit. The volume stored in the circuit never reaches the patient but is instead released through the exhalation valve and is measured with the exhaled gases from the patient. The volume within the ventilator, circuit, and humidifier determines the magnitude of compression volume. The volume loss can theoretically be affected by a number of factors, including temperature and humidity of the circuit, and patient factors such as changes in the patient’s compliance and resistance. In the above-mentioned study, it was found that the reported differences in the measured compliance of the circuit by the manufacturer and those measured by the ventilator with differences were between 37% and 65% for the infant circuit and between 13% and 23% for the adult circuit. Because this factor is collected at startup of the ventilator and includes any compression volume in the ventilator and humidification system, this number will be higher. Subsequently, the compensation measured in the circuit and ventilator system was 51% and 18% higher, respectively, than could be expected from the compression volume calculated using the compliance factor alone. Thus, if only the circuit compliance factor is used, then the volume measurement at expiration, which accounts for both volume of the circuit and volume delivered to the patient, will be higher than the calculated effective volume.

Collectively, studies have shown that the ventilator-displayed VT, without software compensation for circuit compliance, generally overestimates the true delivered VT. Conversely, when the circuit-compliance compensation feature image is on, the ventilator-displayed VT generally underestimates the true delivered VT.


Phase Variables

Discussion to this point has focused on the control variable required for a mechanical ventilator to deliver a breath to the patient and the interactions that occur with the delivery of that breath. The following discussion focuses on what have been described as phase variables (11). Phase variables control the ventilator during the period of time between the beginning of one breath and the initial phase of the next breath. In other words, phase variables are important in determining how a ventilator initiates, sustains, and ends inspiration and what it does between inspirations. Expiration, being passive, is not described in this terminology. In each phase, a particular variable is measured and used to initiate, sustain, and end the phase. The phase variables include the trigger variable (determines the initiation of inspiration), limit variable (determines what sustains inspiration), and cycle variable (determines the termination of inspiration).


Trigger Variables

Patient ventilator system interactions can be initiated under two settings: The ventilator can deliver a controlled breath independent of the patient’s desire, or it can be coordinated with the patient’s effort. Ventilators will measure one or more of the variables associated with the equation of motion (e.g., pressure, volume, flow, or time). Inspiration is initiated when one of these variables reaches a preset variable. A patient-triggered breath, sometimes known as interactive ventilation, provides patients with some autonomy to alter breathing patterns in response to their ventilatory demand. Such systems necessitate an interface between the ventilator and the patient to allow for rapid, measured responses from the ventilator to meet patient needs. For such systems to operate, they must sense a signal from the patient and recognize the beginning of inspiration. Second, they must pressurize the system to allow for delivery of the breath to the patient. Finally, the system must recognize the end of inspiration and thus termination of the breath. Ideally, if this interaction could be facilitated by direct interaction between the patient and mechanical ventilator, delays and patient discomfort created by the temporary or relative unavailability of the caregiver at the patient’s bedside could be eliminated.

Initial recognition of the signal from the patient to begin inspiration is commonly referred to as triggering. Triggering can be subdivided into pretrigger and trigger phases (12). The pretrigger phase has been defined as the time from the onset of inspiration until triggering occurs. The trigger phase is the time
from triggering until the maximum flow of gas occurs. The most common trigger variables are time and flow.

In time triggering, the ventilator initiates a breath according to a set frequency independent of the patient’s spontaneous efforts. In flow triggering, the ventilator senses the patient’s inspiratory effort as a change in flow from the baseline flow and begins inspiration independent of the set breath frequency. Ventilator features that affect the trigger phase include the response time of the ventilator and the presence of bias flow. Bias flow is a continuous delivery of fresh gas circulating through the inspiratory and expiratory limbs of the circuit. Theoretically, bias flow reduces the WOB by making flow available to satisfy the earliest demand of the patient during inspiration, before the flow is initiated during the pretrigger phase. Increased patient effort to trigger the ventilator and delayed response of the ventilator to the patient’s effort can be translated directly into increased WOB.

Current ventilator designs have improved patient-ventilator interactions by improving both the signal sensed by the ventilator and the response time of the ventilator. Today, all ventilators have the capability to utilize a flow signal as the trigger signal from the patient to the ventilator. Flow triggering has the advantage of allowing the patient to trigger the ventilator with less effort, and it has a faster response time (13). The process of creating somewhat seemingly small amounts of negative pressure in pressure-triggered breaths is made increasingly more difficult when the patient has a smaller ETT and/or in the presence of PEEPi.

Theoretically, patient-triggered ventilation could be improved if a signal could be acquired from the patient that represented the earliest attempt by the patient to acquire a breath and if that signal could represent the amount of effort or drive from the patient for that breath. This approach is represented in what has been termed neurally adjusted ventilatory image assist (NAVA) (14). The NAVA approach to mechanical ventilation is based on the acquisition of the patient’s neural respiratory output as it is transmitted through the phrenic nerve to the diaphragm. This signal is acquired via an esophageal catheter with an imbedded series of electrodes that capture the electrical activity signal of the diaphragm, known as Edi. NAVA responds by providing the requested level of ventilatory support to the patient from the Edi. The advantage of this system is the ability to acquire the patient’s desire to trigger the ventilator quickly and to offer feedback between patient effort and ventilator output. At this writing, NAVA is being investigated in clinical trials in Europe in neonatal, pediatric, and adult patients.


Work of Breathing

Work is equal to the force applied to an object multiplied by the distance the object travels. That is, work = force × distance, or W = F × D. If work is applied into the three dimensions of the respiratory system, work becomes the pressure applied to yield a change in the volume of the system and can be expressed as:


where image P is the integral of the pressure across the respiratory system as a function of volume, and dv is the change in the volume of the respiratory system.

The concept of work associated with the functioning of the respiratory system has been known since the seminal analysis of Otis et al. (15). They elucidated that several forces were encountered while breathing, including the elastic forces of the chest wall and lungs, viscous and turbulent resistance of air, nonelastic tissue impedance, and inertia. Basically, motion requires work. Work is performed when pressure changes the volume of the respiratory system and is the product of pressure and volume integrated over time with respect to volume. Work is performed on the respiratory system by externally applied pressures from the ventilator via positive pressure, respiratory muscles, or both, as the lungs expand and contract. To achieve normal ventilation, the body performs work (WOB) to overcome the elastic and frictional resistance of the lungs and chest wall. Total work of breathing (WOBT) is the sum of elastic work (WOBE) and resistive work (WOBR). Elastic WOB represents physiologic work to expand the lungs and chest wall. Resistive WOB is considered a measure of imposed WOB and includes work caused by the breathing apparatus, such as the ETT, breathing circuit, and ventilator demand-flow system. Artificial airways and physiologic resistive work of the airways are responsible for a large part of the imposed resistive work, with the mechanical ventilator also contributing some portion of resistive work (16).

Clinicians have long recognized that increased WOB occurs in patients being weaned from prolonged mechanical ventilation, when the patient begins to breathe spontaneously and take on more of the WOB. Patient-related factors, equipment factors, and decision making affect weaning of patients from mechanical ventilation and, thus, WOB. Equipment factors relate to the ability of the mechanical ventilator to meet the needs of the patient. It has been demonstrated in a lung model that the amount of WOB varies according to the device utilized (17). These equipment factors have an increased significance in patients with poor pulmonary reserve or high airway resistance, where the WOB associated with the equipment is increased (18,19). These factors also have an increased significance in pediatric patients, in whom equipment is often associated with increased WOB (20).


Limit Variable

The limit variable is the modality that sustains inspiration. Inspiration time is defined as the time interval from the beginning of inspiratory flow to the beginning of expiratory flow. During inspiration, pressure, volume, and flow increase above their end-expiratory values.

If one or more of these variables increases only as high as a preset value, this variable will be referred to as the limit variable. It is important to recognize that the limit variable determines what sustains inspiration but differs from the cycle variable, which determines the end of inspiration. Therefore, a limit value does not terminate inspiration but increases it to a preset value.


Cycle Variable

The cycle variable is the modality that terminates inspiration once a preset value is obtained, and this variable must be measured. The cycle variable also differs according to the mode of ventilation used. In pressure-support ventilation (PSV

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Jun 4, 2016 | Posted by in CRITICAL CARE | Comments Off on Mechanical Ventilation

Full access? Get Clinical Tree

Get Clinical Tree app for offline access