Blood pressure is the product of
CO and systemic vascular resistance (
SVR). As with other hemodynamic parameters, blood pressure should be placed in its clinical context and used in conjunction with other monitoring modalities to create a comprehensive hemodynamic profile. Measuring blood pressure may be achieved in several ways; however, due to the frequency required in the critically ill setting, only two are practical: the noninvasive oscillometric method and invasive arterial monitoring.
Noninvasive Oscillometric Arterial Pressure Measurement
Noninvasive blood pressure monitoring is safe, simple, relatively reliable, reproducible, and inexpensive. It should be routinely performed in all children receiving inpatient care. The idea of sphygmomanometry was conceived by von Basch in the 1870s. However, the first indirect method, based on applying a known external pressure to occlude an artery, was developed in the 1890s. A cuff placed around the arm or leg abolishes the pulse when inflated above the patient’s systolic blood pressure. Deflating the cuff slowly allows turbulent blood flow in the artery to develop, which can be appreciated either by auscultation or by observing the oscillations of a manometer reflecting pulsatile flow. Automated oscillometric technology with cuff pressure sensors and a microprocessor controlling the sequence of inflation and incremental deflation is now the standard of care in critical care units. Maximal oscillation amplitude corresponds to the mean arterial pressure. A microprocessor algorithm derives the systolic and diastolic pressure from the change in slope of oscillation amplitude. Oscillation frequency readily yields the heart rate.
The frequency of obtaining blood pressure measurement depends on the severity of disease. Frequent measurements (every 1-3 minutes) are advised during resuscitation or procedures such as intubation. During these occasions care must be taken to ensure that an inflated cuff will not interfere with the administration of drugs from a distal
IV or interrupt pulse oximetry monitoring. Ulnar nerve palsy is a potential complication when pressures are measured frequently for a prolonged period.
Technical difficulties can occur if the patient is small, obese, edematous, agitated (moving), shivering (local muscle contraction causes pseudohypertension), extremely tachycardic, or suffering from burns. Error may arise with an inappropriate size cuff. The cuff should cover at least two thirds of the upper arm, and one too small may overestimate blood pressure and vice versa. Petechial rashes have been noted in the area under the cuff usually reflecting repeated external pressure but may reveal an underlying coagulation defect. Oscillometric blood pressure is not suitable for pressure measurement during nonpulsatile flow such as extracorporeal membrane oxygenation. Finally, the oscillometric-derived blood pressure tends to underestimate or show a lower-than-actual diastolic pressure (
12).
Invasive Arterial Pressure Monitoring
Invasive arterial monitoring is the “gold standard” but is not without risk. Its higher degree of complexity increases the chance of technical errors and complications. However, a number of advantages over its noninvasive counterpart exist: improved accuracy, continuous beat-to-beat measurement, waveform analysis, and frequent arterial blood sampling. The Reverend Stephen Hales first demonstrated invasive blood pressure monitoring in a horse, observing pressure as a column of blood in a glass tube connected to an artery. The principle for today’s measurement systems consists of a column of fluid that directly connects the arterial system to a strain gauge transducer, or Wheatstone bridge, which alters resistance to electron flow with variable pressure. Modern transducers utilize silicon crystals within a semiconductor that change electrical resistance in proportion to applied pressure.
To ensure accuracy and to counteract baseline drift, the transducer must regularly be zeroed to atmospheric pressure at the level of the right atrium (
RA), which corresponds to the mid-axillary line. If the transducer is placed below the reference level, the arterial pressure reading will be falsely elevated and vice versa.
A transmitted arterial pressure waveform is converted into an electrical signal, amplified, and displayed continuously. The arterial pressure waveform represents the summation of a series of sine waves of different frequency, amplitude, and phase. It primarily consists of a fundamental wave (the pulse rate) and a series of further harmonics. Harmonics are smaller waves whose frequencies are multiples of the fundamental frequency. Fourier analysis, described by Lord Kelvin as “a great mathematical poem,” allows the waveform to be examined in terms of its constituent parts then reconstructed and displayed on the bedside monitor in a manner that is simple to interpret.
Monitoring systems are designed to have dynamic response characteristics brisk enough to ensure accurate reproduction of these waveforms across the wide range of heart rates and frequencies encountered in clinical practice. However, every substance has its own natural frequency at which it will oscillate freely. Any component sine wave of the arterial waveform close to that of the monitoring system’s natural (or resonant) frequency will cause resonance, which accentuates the waveform. This is of particular importance in pediatrics, because high heart rates can approach the system’s natural frequency resulting in an exaggerated and distorted signal, falsely elevating the systolic pressure.
In addition to the frequency response, the optimal dynamic capability and accuracy of a system also depends on its damping coefficient. All monitoring systems produce natural energy at rest through oscillation, which can create artifact and therefore possible distortion of the resultant waveform. This is counteracted by an inherent damping capability, dissipating this natural energy by frictional forces in the system. The degree of damping is rarely perfect. Too much (overdamping) or too little (underdamping) may falsely lower or elevate systolic pressures, which, if unrecognized, could influence therapeutic decision-making (
Fig. 71.1). Overdamping is most commonly encountered in the
ICU and occurs with obstruction or excessive compliance in the system resulting in a narrow pulse pressure and a flattened appearance on the displayed waveform. The mean arterial pressure is usually unaffected. Causes include large bubbles; clots; compliant, cracked, lengthy, or kinked tubing; soft transducer diaphragm; three-way taps; or a poorly secured transducer. If this picture emerges then it is worth examining the system for air bubbles and aspirating and flushing the cannula. Smaller-diameter cannula cause overdamping but this cannot be avoided in younger children. Underdamping has the opposite effect. Again, the mean pressure remains largely unchanged. Accurate invasive systems can be achieved by using a short length of wide, stiff tubing, filled with low-viscosity fluid, and free of air bubbles and clot. This ensures that the natural frequency is usually sufficiently high to overcome the problem of resonance and optimal energy dissipation and damping. The system is continuously flushed to reduce the chance of clotting. The ideal solution for this is dextrose due to its nonconducting properties; however; with children, and especially small babies, there is the risk of administering relatively large volumes of free water, which may lead to hyponatremia. Therefore, line patency is maintained by using saline (with or without heparin), which is driven by a 500-mL bag that is pressurized to 300 mm Hg, delivering ˜1-2 mL/h/lumen. A syringe driver (pump) is the
preferred method in smaller children due to improved accuracy of volume infused, which should be taken into account when calculating fluid balance. Infusion sets should be renewed every 72 hours to minimize line infections.
Interpretation
Invasive blood pressure monitoring provides a continuous display of the arterial waveform and additional information not available with the noninvasive modality (
Fig. 71.2). The slope of the upstroke of the arterial waveform may be proportional to myocardial contractility. A slow upstroke can be indicative of poor cardiac function but is also seen in aortic stenosis and elevations in
SVR. The area under the systolic portion of the waveform is proportional to the stroke volume. A low pulse pressure may reflect a low stroke volume. A widened pulse pressure due to elevated systolic and depressed diastolic pressures is seen in states characterized by poor systemic vascular tone, in patients with an aorta-to-pulmonary artery runoff (e.g., patent ductus arteriosus and aortopulmonary artery window), and in patients with aortic insufficiency.