Complications of Nucleus Replacement and Motion-Sparing Technologies

Chapter 9 Complications of Nucleus Replacement and Motion-Sparing Technologies




Chapter Overview


Chapter Synopsis: Spinal fusion surgery can provide successful outcomes in treatment of low back pain arising from disc degeneration, but the surgery can result in spinal degeneration adjacent to the fusion. So-called motion-sparing technologies present newer alternatives. Lumbar arthroplasty replaces the degenerating collagen disc with newly improved artificial materials. Similarly, nucleoplasty replaces the disc’s watery center to preserve structural integrity. These technologies may provide better outcomes and prevent adjacent disc degeneration, but further long-term studies are needed before these more expensive procedures become routine. In addition, the procedures carry many of the same risks as fusion surgery, as well as other potential device-related complications.


Important Points:








Introduction


Surgery for the management of degenerative disorders of the spinal column continues to evolve. Although a majority of these procedures address simple decompression of the neural elements, the use of spinal fusion has been rapidly increasing. Fusion has long been accepted as a means for managing complex spinal deformities and instability. However, the primary reason for its increasing frequency has been its application to the management of degenerative disc disease and the treatment of chronic low back pain. Although this application remains controversial among spinal surgeons, several studies have shown that fusion for degenerative disc disease at one or two levels can improve outcomes compared with the natural history of low back pain.1,2


Although many patients do well after lumbar fusion surgery, some patients can develop further degeneration at spinal levels adjacent to the fusion. Numerous biomechanical studies have demonstrated an increase in mechanical stresses within motion segments adjacent to fused segments.35 As adjacent segment degeneration progresses, additional surgery to decompress and stabilize the involved segments may be necessary. This results in increased morbidity and expense of care and can propagate the cycle of additional adjacent segment degeneration and the subsequent need for even more surgery.


This long-standing problem with adjacent segment degeneration has led many to investigate alternatives to spinal fusion as a means of reducing the mechanical stresses placed on these segments. In particular, the development of a variety of motion-sparing technologies has been enthusiastically embraced as the ideal technology to manage degenerative disorders of the spine while limiting the potential for adjacent segment degeneration.


Motion-sparing technology encompasses a spectrum of devices. The most common of these devices are artificial discs for both the lumbar and cervical regions. Other motion-sparing technologies include nucleus replacement devices, interspinous devices, facet replacement devices, and posterior lumbar dynamic stabilization devices.


Each of these devices allows for some motion at the operated levels, subsequently decreasing the mechanical stresses felt at adjacent levels.6 In theory, this may reduce the incidence of degeneration at adjacent segments caused by fusion, resulting in improved long-term patient outcomes. This technology is attractive to both spinal surgeons as well as informed patients who both recognize the theoretical advantages to a spine fusion alternative. However, these devices are costly, and many lack the long-term follow-up needed to justify the added expense.


Although motion-sparing technologies have been developed to reduce the long-term complications associated with fusion procedures, they are not immune from their own set of postoperative problems. This chapter reviews the complications that can occur through the use of motion-sparing technology, specifically lumbar artificial disc replacement (arthroplasty) and lumbar nucleus replacement (nucleoplasty).



Lumbar Arthroplasty


The search for a surgical alternative to fusion and a means of treating painful lumbar discs dates back several decades. The initial efforts to design and develop an artificial lumbar disc began in the 1950s. Numerous designs were created and patented with few reaching the level of clinical study because of limitations with biomaterial design, lack of appropriate surgical approaches, and inconclusive patient selection strategies.


These early failures with artificial disc technology gradually improved aided by advancements in the design and application of artificial hip and knee joints. In the early 1980s Shellnack and Buttner-Janz7 introduced the SB Charité lumbar artificial disc. The device used a sliding, unconstrained polyethylene core placed between two metallic endplates. When inserted into the affected disc space through an anterior surgical approach, it permitted relatively normal motion of the intervertebral segment.


After two early design changes, the SB Charité disc (DePuy Spine, Raynham, MA) eventually underwent a rigorous multicenter, prospective, randomized study with encouraging early and midterm results. It became the first lumbar disc device to receive Food and Drug Administration (FDA) approval. Widespread use of the device began in 2004. The ProDisc device (Synthes Spine, Paoli, PA) also received early FDA approval and is currently in use (Fig. 9-1). Several other lumbar disc replacement devices are currently in varying stages of design, development, clinical analysis, and approval.





Clinical Results


Lumbar arthroplasty has been promoted as a replacement for fusion. Preserving intersegmental motion is believed to reduce the potential for the development of adjacent level degenerative changes that can be seen after fusion procedures. Several early studies were retrospective in design and demonstrated superior results for disc replacement.8,11,12 David13 evaluated 106 patients who had received implantation of a Charité disc with a minimum 10-year follow-up. In this study, 87 patients (82.1%) achieved excellent or good clinical outcome. Of the 96 patients working preoperatively, 86 (89.6%) returned to work. A total of 90.6% of the implanted devices were still mobile at long-term follow-up.


The first major randomized, controlled clinical study of lumbar arthroplasty evaluated the use of the Charité device. The control group selected for this study was the use of an anterior lumbar interbody fusion (ALIF) with threaded titanium cages. The selection of this control group has been criticized as an outdated technology not in current use because of relatively high failure rates and better alternatives.14 The use of threaded cages before the study was indicated primarily in patients with a collapsed and painful disc space. However, these patients were excluded from the arthroplasty study. Only patients with preserved disc height were included, a suboptimal criterion for the use of threaded cages.


The study defined success as a 25% improvement in Oswestry Disability Index (ODI), no device failure, no major surgical complication, and no neurologic deterioration.15 With the control group selected for this study, it could be predicted that the patients in this group would not do well. This was the case because only 46.5% of the control group patients were classified as achieving success. This is a poor result compared with other series of ALIF in properly selected patients in which success rates in the 85% to 95% range have been reported.16,17


The arthroplasty group did only slightly better achieving a success rate in only 57.1% of the patients. This modest success rate is further limited by the fact that at 2 years of follow-up, 72.2% of the arthroplasty patients were still on narcotic medication. However, the study did demonstrate safety and effectiveness of the device and ultimately contributed to final FDA approval.15


Currently, several other implant designs are in various stages of clinical investigation. The two devices approved for clinical use are gathering longer term follow-up in an effort to determine the effectiveness of arthroplasty at limiting adjacent segment degeneration. To date, little evidence exists to support a significant reduction in adjacent segment degeneration with the use of arthroplasty. Until sufficient long-term follow-up is obtained, reimbursement for the procedure will remain a challenge, further limiting the widespread application of this technology.